Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: RF MultiGen, 0406-900-000; Refurb RF MultiGen, 0406-900-000U

Recall number
Z-0584-2017
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
RF MultiGen, 0406-900-000; Refurb RF MultiGen, 0406-900-000U
Status
Terminated
Initiated
2016-10-11
Posted
2016-11-16
Terminated
2017-07-25
Reason
Stryker Instruments is voluntarily recalling the Care Instructions/Instructions for Use packaged with the Radiofrequency (RF) MultiGen Generator, because the instructions for use incorrectly contained instructions supporting the sterilization of the MultiGen cables.
Root cause
Labeling Change Control
Action
Stryker Instruments initiated a voluntary recall of the RF MultiGen, and Refurb RF MultiGen because the instructions for use contained incorrect instructions on sterilizing the device. Stryker distributed customer notices via certified mail on 10/11/2016. Customer's were instructed to do the following: Immediately review this recall notification. 2. Enclosed was a revised version of the IVS Care Instructions. Please distribute the revised version to the appropriate person or department. 3. Remove and discard any previously distributed IVS Care Instructions or Instructions for Use. 4. Please complete, sign, and return the enclosed Business Reply Form (BRF) acknowledging your receipt and understanding of this notification. 5. Fax the completed BRF to Stryker Instruments Regulatory Department, 866-521-2762, or scan and email a copy to kellyjo.davis@stryker.com. Customers with questions can contact: 269 323 7700 .
Quantity
2,675
Distribution
Domestic:AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign:France, Germany, Canada, UK, Lebanon, South Africa, Chile, Hong Kong, Poland, Argentina, Spain, Greece, Italy, India, Brazil, Switzerland, Netherlands, Sweden VA/DOD:
Lot, serial or model codes
RF MultiGen¿, 0406-900-000, Refurb RF MultiGen¿, 0406-900-000U
510(k) numbers
["K071482"]
PMA numbers
not on file
Product code
GXD
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register