FDA Medical Device Recalls (openFDA)
Smith & Nephew, Inc.: BIRMINGHAM HIP RESURFACING MODULAR HEAD, REF numbers: 74121238, MODULAR HEAD 38 MM -8MM 74121242, MODULAR HEAD 42 MM -8MM 74121246, MODULAR HEAD 46 MM -8MM 74121250, MODULAR HEAD 50 MM -8MM 74121254, MODULAR HEAD 54 MM -8MM 74121258, MODULAR HEAD 58 MM -8MM 74121338, MODULAR HEAD 3 8MM -4 (SHORT) 74121342, MODULAR HEAD 42 MM -4 (SHORT) 74121346, MODULAR HEAD 46 MM -4 (SHORT) 74121350, MODULAR HEAD 50 MM -4 (SHORT) 74121354, MODULAR HEAD 54 MM -4 (SHORT) 74121358, MODULAR HEAD 58 MM -4 (SHORT) 74121438, MODULAR HEAD 38 MM +0 (MED) 74121442, MODULAR HEAD 42 MM +0 (MED) 74121446, MODULAR HEAD 46 MM +0 (MED) 74121450, MODULAR HEAD 50 MM +0 (MED) 74121454, MODULAR HEAD 54 MM +0 (MED) 74121458, MODULAR HEAD 58 MM +0 (MED) 74121538, MODULAR HEAD 38 MM +4 (LONG) 74121542, MODULAR HEAD 42 MM +4 (LONG) 74121546, MODULAR HEAD 46 MM +4 (LONG) 74121550, MODULAR HEAD 50 MM +4 (LONG) 74121554, MODULAR HEAD 54 MM +4 (LONG) 74121558, MODULAR HEAD 58 MM +4 (LONG)
- Recall number
- Z-0584-2016
- Recalling firm
- Smith & Nephew, Inc.
- Product
- BIRMINGHAM HIP RESURFACING MODULAR HEAD, REF numbers: 74121238, MODULAR HEAD 38 MM -8MM 74121242, MODULAR HEAD 42 MM -8MM 74121246, MODULAR HEAD 46 MM -8MM 74121250, MODULAR HEAD 50 MM -8MM 74121254, MODULAR HEAD 54 MM -8MM 74121258, MODULAR HEAD 58 MM -8MM 74121338, MODULAR HEAD 3 8MM -4 (SHORT) 74121342, MODULAR HEAD 42 MM -4 (SHORT) 74121346, MODULAR HEAD 46 MM -4 (SHORT) 74121350, MODULAR HEAD 50 MM -4 (SHORT) 74121354, MODULAR HEAD 54 MM -4 (SHORT) 74121358, MODULAR HEAD 58 MM -4 (SHORT) 74121438, MODULAR HEAD 38 MM +0 (MED) 74121442, MODULAR HEAD 42 MM +0 (MED) 74121446, MODULAR HEAD 46 MM +0 (MED) 74121450, MODULAR HEAD 50 MM +0 (MED) 74121454, MODULAR HEAD 54 MM +0 (MED) 74121458, MODULAR HEAD 58 MM +0 (MED) 74121538, MODULAR HEAD 38 MM +4 (LONG) 74121542, MODULAR HEAD 42 MM +4 (LONG) 74121546, MODULAR HEAD 46 MM +4 (LONG) 74121550, MODULAR HEAD 50 MM +4 (LONG) 74121554, MODULAR HEAD 54 MM +4 (LONG) 74121558, MODULAR HEAD 58 MM +4 (LONG)
- Status
- Terminated
- Initiated
- 2015-11-16
- Posted
- not on file
- Terminated
- 2016-08-25
- Reason
- The 16 devices distributed in the United States were incorrectly labeled as the monoblock Birmingham Hip Modular Head, with labeling not cleared for marketing in the US. The actual devices were monoblock Modular Femoral Head for hemiarthroplasty use.
- Root cause
- Labeling design
- Action
- The firm sent their field safety notices on 11/16/2015 by letter delivered by Fed Ex (overnight) and email. Non- responding consignees will be contacted within 4 weeks of the initial notice
- Quantity
- 16 units in US
- Distribution
- Worldwide distribution. US nationwide, France, UAE, Austria, Australia, Belgium, CANADA, Switzerland, Germany, Denmark, Spain, Finland, UK, Hong Kong, ISRAEL, India, Italy, Japan, South Korea, Netherlands, Norway, Poland, PORTUGAL, Sweden, SINGAPORE, Thailand, Turkey, and SOUTH AFRICA.
- Lot, serial or model codes
- All codes
- 510(k) numbers
- ["K061243"]
- PMA numbers
- not on file
- Product code
- KWL
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28