Skip to the content

FDA Medical Device Recalls (openFDA)

DERMASENSOR INC: DermaSensor. Model/Catalog Number: 10101. Software Version: 3.0.2. Software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer

Recall number
Z-0583-2026
Recalling firm
DERMASENSOR INC
Product
DermaSensor. Model/Catalog Number: 10101. Software Version: 3.0.2. Software-aided adjunctive diagnostic device for use by physicians on lesions suspicious for skin cancer
Status
Open, Classified
Initiated
2025-10-13
Posted
2025-11-21
Terminated
not on file
Reason
Potential for device to not meet specifications resulting in incorrect results or delay of referral for patients.
Root cause
Process control
Action
DermaSensor began notification of consignees on about 10/13/2025 via telephone call with a follow up letter. The telephone call notified consignees of the issue and confirmed if the affected units were or were not on hand. The letter then sent to the consignees again advised them of the issue and instructed them to examine inventory for affected units, to complete and return the Customer Notification Response Form, and arrange a visit with DermaSensor for the device correction. Once the affected unit is examined, DermaSensor can determine if the device contains the component issue and if it does, DermaSensor will provide the consignee with the dates the scans were taken and the correct results. If the consignee's device does not contain the component issue, no further actions need to be taken.
Quantity
343 (9 units affected)
Distribution
US distribution to MD, FL, AZ, MS, PA, NY, CT, CA.
Lot, serial or model codes
Lot Code: Model No 10101; UDI-DI 085001778610101A9; Serial Number 1010120241230196E4E, 1010120241240151FD5, 101012024130014EF69, 101012024125014F595, 10101202413001F7C66, 1010120242400172717, 101012024325014CE26, 10101202433101D295C, 101012025093012EE03.
510(k) numbers
not on file
PMA numbers
not on file
Product code
QZS
Firm FEI
3017254114
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register