Baxter Healthcare Renal Div: Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
- Recall number
- Z-0583-04
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
- Status
- Terminated
- Initiated
- 2004-02-20
- Posted
- 2004-07-20
- Terminated
- 2008-02-26
- Reason
- These devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.
- Root cause
- Other
- Action
- Baxter sent urgent device correction letters dated 2/20/04 to the Hemodialysis Unit Administrators on the same date. The letters informed the accounts of the potential of a heater system failure that under specific conditions could lead to a fire. The letters listed the concurrent conditions that must exist for the potential overheating, and recommended that the following steps be taken to ensure the safety in operating the devices: 1) Plug these machine into GFCI outlets if available, and 2) If GFCI outlets are not available at the facility, the machines must be 'hard-powered off' after dialysis treatments are complete at the end of the day. The machine is had-powered off by placing the main power switch located in the back of the machine into the OFF position. If the facility typically performs a heat clean cycle of machines after treatments are completed, it is imperative that the machines are hard powered off at the completion of the cycle. Any questions were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 1, Monday through Friday, 8:30 AM to 6:00 PM EST.
- Quantity
- 28,482 units
- Distribution
- The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Lebanon, Morocco, Malaysia, the Netherlands, Oman, Panama, Preu, Philippines, Poland, Palestine, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and United Kingdom.
- Lot, serial or model codes
- Serial numbers 1001S to 1142S; 50001 to 52109; 01001 to 26231
- 510(k) numbers
- ["K970446"]
- PMA numbers
- not on file
- Product code
- FII
- Firm FEI
- 3000210122
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28