Baxter Healthcare Renal Div: Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

Recall number
Z-0583-04
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
Status
Terminated
Initiated
2004-02-20
Posted
2004-07-20
Terminated
2008-02-26
Reason
These devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.
Root cause
Other
Action
Baxter sent urgent device correction letters dated 2/20/04 to the Hemodialysis Unit Administrators on the same date. The letters informed the accounts of the potential of a heater system failure that under specific conditions could lead to a fire. The letters listed the concurrent conditions that must exist for the potential overheating, and recommended that the following steps be taken to ensure the safety in operating the devices: 1) Plug these machine into GFCI outlets if available, and 2) If GFCI outlets are not available at the facility, the machines must be 'hard-powered off' after dialysis treatments are complete at the end of the day. The machine is had-powered off by placing the main power switch located in the back of the machine into the OFF position. If the facility typically performs a heat clean cycle of machines after treatments are completed, it is imperative that the machines are hard powered off at the completion of the cycle. Any questions were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 1, Monday through Friday, 8:30 AM to 6:00 PM EST.
Quantity
28,482 units
Distribution
The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Lebanon, Morocco, Malaysia, the Netherlands, Oman, Panama, Preu, Philippines, Poland, Palestine, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and United Kingdom.
Lot, serial or model codes
Serial numbers 1001S to 1142S; 50001 to 52109; 01001 to 26231
510(k) numbers
["K970446"]
PMA numbers
not on file
Product code
FII
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register