Skip to the content

FDA Medical Device Recalls (openFDA)

Skeletal Dynamics: Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: Joint destruction and/or subluxation and resistance to conservative treatment. Intended use is also for primary replacement after fracture of the radial head and symptomatic sequelae after radial head resection.

Recall number
Z-0582-2015
Recalling firm
Skeletal Dynamics
Product
Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: Joint destruction and/or subluxation and resistance to conservative treatment. Intended use is also for primary replacement after fracture of the radial head and symptomatic sequelae after radial head resection.
Status
Terminated
Initiated
2014-10-16
Posted
2014-12-12
Terminated
2016-12-14
Reason
Report received where the Align Radial Stem fractured.
Root cause
Device Design
Action
Skeletal Dynamic sent an Urgent Medical Device Safety Alert dated October 16, 2014, to all affected customers via certified mail. The Safety Alert included the warning "that radial head prostheses cannot be expected to withstand the activity levels and loads of normal healthy bone and joint tissue. Failure of the component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union or excessive loads (estimated body weight equivalent of 350 lbs or greater)" with instructions to notify any surgeons that have used the product. A response form has been included with instructions to return the signed form to acknowledge the receipt and understanding of the new warning. Product labeling has been revised to include these labels in future shipments. Customers with questions were instructed to contact Customer Service at 877-753-5396 for assistance. For questions regarding this recall call 305-596-7585, ext 7026.
Quantity
1700
Distribution
Nationwide Distribution including GA, NJ, NM, IL, CA, OK, CO, NY, LA, OH, FL, VA, MD, MS, TX, IA, AL, WA, TN, ME, CT, WI, AZ, NE, SC, NC and Puerto Rico.
Lot, serial or model codes
Cat #s: ALN-RST-0700, ALN-RST-0702, ALN-RST-0704, ALN-RST-0706, ALN-RST-0708, ALN-RST-0800, ALN-RST-0802, ALN-RST-0802, ALN-RST-0804, ALN-RST-0806, ALN-RST-0808, ALN-RST-0900, ALN-RST-0902, ALN-RST-0904, ALN-RST-0906, ALN-RST-0908, ALN-RST-1000, ALN-RST-1002,  ALN-RST-1004, ALN-RST-1006, ALN-RST-1008, ALN-RST-1100, ALN-RST-1102, ALN-RST-1104, ALN-RST-1106, ALN-RST-1108; Lot #s: AN0610023, AN0212065, AN0913033, AN0913034, AN0610024, AN0913035, AN0913036, AN0610025, AN0913037, AN0913038, AN0610026,  AN0913039, AN0913040, AN0610027, AN0913041, AN0913042, AN0610028, AN0212066, AN0913043,  AN0913044, AN0610029, AN0913045, AN0913046,  AN0610030, AN0913047, AN0913048, AN0610031,  AN0913049, AN0913050, AN0610032, AN0913051,  AN0913052, AN0610033, AN0212067, AN0913053, AN0913054, AN0610034, AN0913055, AN0913056, AN0610035, AN0913057, AN0913058, AN0610036,  AN0913059, AN0913060, AN0610037, AN0913061,  AN0913062, AN0610038, AN0212068, AN0913063, AN0913064, AN0610039, AN0212069, AN0913065, AN0913066, AN0610040, AN0913067, AN0913068,  AN0610041, AN0913069, AN0913070, AN0410008,  AN0610042, AN0410009, AN0913071, AN0913072,  AN0610043, AN0212070, AN0913073, AN0913074,  AN0610044, AN0913075, AN0913076, AN0610045, AN0913077, AN0913078, AN0610046, AN0913079, AN0913080, AN0610047, AN0913081, AN0913082.
510(k) numbers
["K092721"]
PMA numbers
not on file
Product code
KWI
Firm FEI
3006742481
Firm city
Miami
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register