Skip to the content

FDA Medical Device Recalls (openFDA)

Varian Medical Systems: EPROM and Control Software (embedded), version 5.08, for the GammaMed 12i/t Afterloader, manufactured by Varian Medical Systems, Inc, Palo Alto, CA 94304.

Recall number
Z-0582-04
Recalling firm
Varian Medical Systems
Product
EPROM and Control Software (embedded), version 5.08, for the GammaMed 12i/t Afterloader, manufactured by Varian Medical Systems, Inc, Palo Alto, CA 94304.
Status
Terminated
Initiated
2004-02-06
Posted
2004-07-20
Terminated
2004-09-21
Reason
Radiation treatment software could potentially cause a healthcare practitioner to apply incorrect treatment regimen to patients undergoing brachytherapy.
Root cause
Other
Action
The recalling firm notified end users by letters flagged 'Urgent Medical Device Correction' on 2/2/04. The notification reminds users to verify the decay factor before patient treatment. The letter also advises that instrument software be be updated commencing March 2004.
Quantity
203
Distribution
Product was distributed to end users in the United States and internationally.
Lot, serial or model codes
Serial numbers (Domestic units): 724 215 765 210 212 217 218 214 767 213 743 706708 784 727 795 710 899 702 898 820 219 750 757 894 799 852 821 216 788 202 873 786 896 797 826 709 889 848
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAQ
Firm FEI
3000206172
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register