Natus Neurology DBA Excel Tech., Ltd. (XLTEK): Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks /SleepWorks software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals.

Recall number
Z-0581-2016
Recalling firm
Natus Neurology DBA Excel Tech., Ltd. (XLTEK)
Product
Natus Quantum System with NeuroWorks Software. Catalog /Part Numbers: 013926. The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks /SleepWorks software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals.
Status
Terminated
Initiated
2015-11-11
Posted
not on file
Terminated
2016-10-03
Reason
During an internal testing, Natus Medical Incorporated, identified that in Neuro Works 8.1 with Quantum hardware, the incoming signal is displayed with reversed polarity.
Root cause
Software Manufacturing/Software Deployment
Action
The firm, Natus Neurology, sent an "Urgent Medical Device Correction Notification" letter on 11/11/2015. The letter described the the problem, the product involved in the recall and actions to be taken. Natus will provide further information once an updated software is released. The consignees were instructed to complete and return the confirmation form by mail to Senior Manager Quality and Regulatory Affairs, Excel-Tech, Ltd (XLTEK) Division, Natus Medical Incorporated, 2568 Bristol Circle, Oakville, ON L6H 5S1 or by email to :ots@natus.com. If you have any questions, contact Senior Manager Quality and Regulatory Affairs at 905-287-5055 or email: sanjay.mehta@natus.com.
Quantity
56 (43 US, 13 OUS)
Distribution
Worldwide Distribution: US Distribution including states of: AL, AZ, AR, CO, FL, GA, MD, MA, MN, MS, NJ, NY, OH, SC, TN, TX and UT; and countries of: INDIA, KOREA, SPAIN, SWITZERLAND, THAILAND, UNITED KINGDOM, and CANADA. .
Lot, serial or model codes
Serial Number 013926020008 013926020009 013926020006 013926020016 013926020014 013926030024 013926030031 013926020001 013926020004 013926030059 013926030060 013926030074 013926030075 013926030052 013926030017 013926030018 013926030021 013926030053 013926030101 013926030058 013926030019 013926030038 013926030039 013926030044 013926020015 013926030097 013926020013 013926030055 013926030056 013926030081 013926030096 013926030078 013926030048 013926030057 013926030064 013926030047 013926030027 013926030025 013926030042 013926030023 013926030030 013926030036 013926030026 013926030050 013926030080 013926030099 013926030035 013926030046 013926030049 013926030073 013926030020 013926030092 013926030040 013926030041 013926030061 013926030062
510(k) numbers
["K143440"]
PMA numbers
not on file
Product code
GWQ
Firm FEI
3002809369
Firm city
Oakville
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register