Greatbatch Medical: Enpath Myopore Sutureless Myocardial Pacing Lead, Ref: 511211. the contents of the inner package have been sterilized by ethylene-oxide gas. Manufactured by: Enpath Medical, Inc., 2300 Berkshire Lane North, Minneapolis, MN 55441 USA.
- Recall number
- Z-0581-2010
- Recalling firm
- Greatbatch Medical
- Product
- Enpath Myopore Sutureless Myocardial Pacing Lead, Ref: 511211. the contents of the inner package have been sterilized by ethylene-oxide gas. Manufactured by: Enpath Medical, Inc., 2300 Berkshire Lane North, Minneapolis, MN 55441 USA.
- Status
- Terminated
- Initiated
- 2009-09-01
- Posted
- 2009-12-29
- Terminated
- 2011-12-17
- Reason
- The silicone tubing of affected leads was processed outside of validated parameters. Although the material remains biocompatible, test data regarding the performance characteristics are not available.
- Root cause
- Component change control
- Action
- Customers were contacted by phone beginning 09/01/09. They were asked to hold product and determine if they had distributed any product. A written notification followed which identified the product, described the issue and asked for product to be returned. Product was distributed to two consignees and contained the following day. Questions are directed to the company at 612-559-2613.
- Quantity
- 35 units
- Distribution
- CA and MN
- Lot, serial or model codes
- Serial Numbers: 148399, 148400, 148401, 148402, 148403, 148404, 148405, 148406, 148407, 148408, 148409, 148410, 148411, 148412, 148413, 148414, 148415, 148416, 148417, 148418, 148419, 148420, 148421, 148422, 148423, 148424, 148427, 148430, 148431, 148433, 148434, 148435, 148436, 148437, 148438.
- 510(k) numbers
- ["K925099"]
- PMA numbers
- not on file
- Product code
- DTB
- Firm FEI
- 2183787
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28