FDA Medical Device Recalls (openFDA)
Boston Scientific Corp: IntellaMap Orion High Resolution Mapping Catheter; Material number: M004RC64S0; Catalog number RC64S; Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
- Recall number
- Z-0580-2015
- Recalling firm
- Boston Scientific Corp
- Product
- IntellaMap Orion High Resolution Mapping Catheter; Material number: M004RC64S0; Catalog number RC64S; Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
- Status
- Terminated
- Initiated
- 2014-11-20
- Posted
- 2014-12-10
- Terminated
- 2016-08-11
- Reason
- Some IntellaMap Orion High Resolution Mapping Catheters are exhibiting a failure mode in which a glue bond on the Deployment Interlock Button has been observed to fail in the field.
- Root cause
- Device Design
- Action
- Urgent Medical Device Recall Correction Notification letters were sent on November 19, 2014 by overnight delivery. The letter identified the affected product and the problem. Customers were asked to read the Notification completely and ensure that all users are aware. The letter provides users with steps to mitigate the risk in the event that the identified problem occurs. Customers were also asked to complete and return the Customer Acknowledgement Form. A local sales representative can answer questions regarding the Field Correction.
- Quantity
- 184 units
- Distribution
- Worldwide Distribution -- US, including the states of MA, AL, NC, and OK; and, the countries of Germany, France, and Great Britain.
- Lot, serial or model codes
- Lot numbers: 17075550, 17075981, 17083913, 17088826, 17094081, 17101668, 17208445, 17218467, 17251661, 17258617, 17274937, 17292792, and 17308586. Expiry Dates from June 25, 2015 to September 24, 2015.
- 510(k) numbers
- ["K122461"]
- PMA numbers
- not on file
- Product code
- DRF
- Firm FEI
- 3001236349
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28