Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corp: IntellaMap Orion High Resolution Mapping Catheter; Material number: M004RC64S0; Catalog number RC64S; Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.

Recall number
Z-0580-2015
Recalling firm
Boston Scientific Corp
Product
IntellaMap Orion High Resolution Mapping Catheter; Material number: M004RC64S0; Catalog number RC64S; Is indicated for electrophysiological mapping (recording or stimulating only) of the cardiac structures of the heart.
Status
Terminated
Initiated
2014-11-20
Posted
2014-12-10
Terminated
2016-08-11
Reason
Some IntellaMap Orion High Resolution Mapping Catheters are exhibiting a failure mode in which a glue bond on the Deployment Interlock Button has been observed to fail in the field.
Root cause
Device Design
Action
Urgent Medical Device Recall Correction Notification letters were sent on November 19, 2014 by overnight delivery. The letter identified the affected product and the problem. Customers were asked to read the Notification completely and ensure that all users are aware. The letter provides users with steps to mitigate the risk in the event that the identified problem occurs. Customers were also asked to complete and return the Customer Acknowledgement Form. A local sales representative can answer questions regarding the Field Correction.
Quantity
184 units
Distribution
Worldwide Distribution -- US, including the states of MA, AL, NC, and OK; and, the countries of Germany, France, and Great Britain.
Lot, serial or model codes
Lot numbers: 17075550, 17075981, 17083913, 17088826, 17094081, 17101668, 17208445, 17218467, 17251661, 17258617, 17274937, 17292792, and 17308586. Expiry Dates from June 25, 2015 to September 24, 2015.
510(k) numbers
["K122461"]
PMA numbers
not on file
Product code
DRF
Firm FEI
3001236349
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register