Skip to the content

FDA Medical Device Recalls (openFDA)

Orthoscan, Inc.: Ziehm Vision RFD. Interventional fluoroscopic x-ray system

Recall number
Z-0578-2025
Recalling firm
Orthoscan, Inc.
Product
Ziehm Vision RFD. Interventional fluoroscopic x-ray system
Status
Open, Classified
Initiated
2024-08-08
Posted
2024-12-02
Terminated
not on file
Reason
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
Root cause
Radiation Control for Health and Safety Act
Action
Ziehm Imaging, Inc. intends to notify the affected customers directly in the following manner: 1) Letter with return receipt requested, envelope, return, and reply to questionnaire, or E-mail contact. 2) Direct phone calls. Once contact is made and confirmed Ziehm Imaging Inc, shall provide replacement hand switch and instructions to the end user. Wired Hand switch 88036 Vl.02 will be requested to be returned by the customer/end user to Ziehm Imaging Inc. and will be subjected to be scrapped.
Quantity
30
Distribution
U.S. (including Puerto Rico).
Lot, serial or model codes
UDI: EZIEZIEHMVISIONRFD1. Serial Numbers: 20048, 20240, 23805, 21350, 23249, 23250, 23251, 23252, 23253, 23324, 23375, 23376, 23377, 23378, 23379, 23380, 23381, 23382, 23412, 23452, 23460, 23461, 23462, 23463, 23506, 23536, 23721, 23737, 23738, 23739.
510(k) numbers
["K240099"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3005254598
Firm city
Scottsdale
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register