Boston Scientific Corporation: Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493919308200 EMERGE MR, OUS 2.00mm X 8mm; 2. REF/Catalog No. H7493918912200, EMERGE MR US 2.00MM X 12MM; 3. REF/Catalog No. H7493919312200, EMERGE MR, OUS 2.00mm X 12mm; 4. REF/Catalog No. H7493918912250, EMERGE MR US 2.50MM X 12MM; 5. REF/Catalog No. H7493919312250, EMERGE MR, OUS 2.50mm X 12mm; 6. REF/Catalog No. H7493918912300, EMERGE MR US 3.00MM X 12MM; 7. REF/Catalog No. H7493919312300, EMERGE MR, OUS 3.00mm X 12mm; 8. REF/Catalog No. H7493918915150, EMERGE MR US 1.50MM X 15MM; 9. REF/Catalog No. H7493919315150, EMERGE MR, OUS 1.50mm X 15mm; 10. REF/Catalog No. H7493918915200, EMERGE MR US 2.00MM X 15MM, 11. REF/Catalog No. H7493919315200, EMERGE MR, OUS 2.00mm X 15mm; 12. REF/Catalog No. H7493919315250, EMERGE MR, OUS 2.50mm X 15mm; 13. REF/Catalog No. H7493918915300, EMERGE MR US 3.00MM X 15MM; 14. REF/Catalog No. H7493919315300, EMERGE MR, OUS 3.00mm X 15mm; 15. REF/Catalog No. H7493918915350, EMERGE MR US 3.50MM X 15MM; 16. REF/Catalog No. H7493919315350, EMERGE MR, OUS 3.50mm X 15mm; 17. REF/Catalog No. H7493919320200, EMERGE MR, OUS 2.00mm X 20mm; 18. REF/Catalog No. H7493919320250, EMERGE MR, OUS 2.50mm X 20mm; 19. REF/Catalog No. H7493919320300, EMERGE MR, OUS 3.00mm X 20mm; 20. REF/Catalog No. H7493919415120, EMERGE PUSH MR, OUS 1.20mm X 15mm.
- Recall number
- Z-0577-2020
- Recalling firm
- Boston Scientific Corporation
- Product
- Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493919308200 EMERGE MR, OUS 2.00mm X 8mm; 2. REF/Catalog No. H7493918912200, EMERGE MR US 2.00MM X 12MM; 3. REF/Catalog No. H7493919312200, EMERGE MR, OUS 2.00mm X 12mm; 4. REF/Catalog No. H7493918912250, EMERGE MR US 2.50MM X 12MM; 5. REF/Catalog No. H7493919312250, EMERGE MR, OUS 2.50mm X 12mm; 6. REF/Catalog No. H7493918912300, EMERGE MR US 3.00MM X 12MM; 7. REF/Catalog No. H7493919312300, EMERGE MR, OUS 3.00mm X 12mm; 8. REF/Catalog No. H7493918915150, EMERGE MR US 1.50MM X 15MM; 9. REF/Catalog No. H7493919315150, EMERGE MR, OUS 1.50mm X 15mm; 10. REF/Catalog No. H7493918915200, EMERGE MR US 2.00MM X 15MM, 11. REF/Catalog No. H7493919315200, EMERGE MR, OUS 2.00mm X 15mm; 12. REF/Catalog No. H7493919315250, EMERGE MR, OUS 2.50mm X 15mm; 13. REF/Catalog No. H7493918915300, EMERGE MR US 3.00MM X 15MM; 14. REF/Catalog No. H7493919315300, EMERGE MR, OUS 3.00mm X 15mm; 15. REF/Catalog No. H7493918915350, EMERGE MR US 3.50MM X 15MM; 16. REF/Catalog No. H7493919315350, EMERGE MR, OUS 3.50mm X 15mm; 17. REF/Catalog No. H7493919320200, EMERGE MR, OUS 2.00mm X 20mm; 18. REF/Catalog No. H7493919320250, EMERGE MR, OUS 2.50mm X 20mm; 19. REF/Catalog No. H7493919320300, EMERGE MR, OUS 3.00mm X 20mm; 20. REF/Catalog No. H7493919415120, EMERGE PUSH MR, OUS 1.20mm X 15mm.
- Status
- Terminated
- Initiated
- 2019-10-24
- Posted
- not on file
- Terminated
- 2023-03-20
- Reason
- Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
- Root cause
- Process control
- Action
- The firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: BSCFieldActionCenter@bsci.com OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases even if you do not have any affected product. If you are a distributor, please note that the removal notification should be forwarded to your customers. If you have any questions, contact Boston Scientific Quality Systems at 763-494-1133 or email: BSCFieldActionCenter@bsci.com.
- Quantity
- 14,145 devices total
- Distribution
- Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
- Lot, serial or model codes
- Lot / GTIN Number / Use By: 1. 24112923 / 8714729799788 / 2022-03-12 2. Lot: 21362152 / 8714729806080 / 2020-07-06, 3. Lots: 21311605 / 8714729799894 / 2020-06-23, 23960557 / 8714729799894 / 2022-02-11, 23975925 / 8714729799894 / 2022-02-13, 24112926 / 8714729799894 / 2022-03-12. 4. Lots: 20524227 / 8714729806103 / 2019-12-31, 22402212 / 8714729806103 / 2021-03-15, 22909685 / 8714729806103 / 2020-12-06. 5. Lots: 23881230 / 8714729799917 / 2022-01-25, 23881291 / 8714729799917 / 2022-01-25, 24088571 / 8714729799917 / 2022-03-07, 24143068 / 8714729799917 / 2022-03-18. 6. Lots: 22374160 / 8714729806127 / 2021-03-09, 22532219 / 8714729806127 / 2021-04-16, 22760617 / 8714729806127 / 2021-02-05, 23021557 / 8714729806127 / 2021-01-11, 23181062 / 8714729806127 / 2021-09-10, 23578933 / 8714729806127 / 2021-04-16. 7. Lot: 22799195 / 8714729799931 / 2021-06-12. 8. Lots: 20361236 / 8714729806189 / 2019-10-31, 22619082 / 8714729806189 / 2021-05-03. 9. Lot: 24088574 / 8714729799993 / 2022-03-07. 10. Lots: 20985435 / 8714729806196 / 2020-04-04, 22416559 / 8714729806196 / 2021-03-19. 11. Lots: 21308852 / 8714729800002 / 2020-06-23, 21309387 / 8714729800002 / 2020-06-23, 21363021 / 8714729800002 / 2020-07-06, 23989173 / 8714729800002 / 2022-02-15. 12. Lots: 20909603 / 8714729800026 / 2020-03-15, 22734757 / 8714729800026 / 2021-05-29, 23047327 / 8714729800026 / 2021-08-03, 23946181 / 8714729800026 / 222-02-07, 24151134 / 8714729800026 / 2022-03-19. 13. Lot: 23924000 / 8714729806233 / 2022-02-04. 14. Lots: 23272475 /8714729800040 / 2021-09-27 , 23541254 / 8714729800040 / 2021-11-20. 15. Lot: 23901505 / 8714729806257 / 2022-01-30. 16. Lot: 24133473 / 8714729800064 / 2022-03-15. 17. Lots: 23272477 / 8714729800118 / 2021-09-21, 23324727 / 8714729800118 / 2021-10-08. 18. Lots: 20452512 / 8714729800132 / 2019-11-30, 20669615 / 8714729800132 / 2020-01-15, 23408005 / 8714729800132 / 2021-10-24. 19. Lots: 21500615 / 8714729800156 / 2020-08-08, 23518523 / 8714729800156 / 2021-11-15. 20. Lot: 23364688 / 8714729799702 / 2021-10-16.
- 510(k) numbers
- ["K113220","K121196","K130391","K163174"]
- PMA numbers
- not on file
- Product code
- LOX
- Firm FEI
- 3002095335
- Firm city
- Maple Grove
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28