Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493919308200 EMERGE MR, OUS 2.00mm X 8mm; 2. REF/Catalog No. H7493918912200, EMERGE MR US 2.00MM X 12MM; 3. REF/Catalog No. H7493919312200, EMERGE MR, OUS 2.00mm X 12mm; 4. REF/Catalog No. H7493918912250, EMERGE MR US 2.50MM X 12MM; 5. REF/Catalog No. H7493919312250, EMERGE MR, OUS 2.50mm X 12mm; 6. REF/Catalog No. H7493918912300, EMERGE MR US 3.00MM X 12MM; 7. REF/Catalog No. H7493919312300, EMERGE MR, OUS 3.00mm X 12mm; 8. REF/Catalog No. H7493918915150, EMERGE MR US 1.50MM X 15MM; 9. REF/Catalog No. H7493919315150, EMERGE MR, OUS 1.50mm X 15mm; 10. REF/Catalog No. H7493918915200, EMERGE MR US 2.00MM X 15MM, 11. REF/Catalog No. H7493919315200, EMERGE MR, OUS 2.00mm X 15mm; 12. REF/Catalog No. H7493919315250, EMERGE MR, OUS 2.50mm X 15mm; 13. REF/Catalog No. H7493918915300, EMERGE MR US 3.00MM X 15MM; 14. REF/Catalog No. H7493919315300, EMERGE MR, OUS 3.00mm X 15mm; 15. REF/Catalog No. H7493918915350, EMERGE MR US 3.50MM X 15MM; 16. REF/Catalog No. H7493919315350, EMERGE MR, OUS 3.50mm X 15mm; 17. REF/Catalog No. H7493919320200, EMERGE MR, OUS 2.00mm X 20mm; 18. REF/Catalog No. H7493919320250, EMERGE MR, OUS 2.50mm X 20mm; 19. REF/Catalog No. H7493919320300, EMERGE MR, OUS 3.00mm X 20mm; 20. REF/Catalog No. H7493919415120, EMERGE PUSH MR, OUS 1.20mm X 15mm.

Recall number
Z-0577-2020
Recalling firm
Boston Scientific Corporation
Product
Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493919308200 EMERGE MR, OUS 2.00mm X 8mm; 2. REF/Catalog No. H7493918912200, EMERGE MR US 2.00MM X 12MM; 3. REF/Catalog No. H7493919312200, EMERGE MR, OUS 2.00mm X 12mm; 4. REF/Catalog No. H7493918912250, EMERGE MR US 2.50MM X 12MM; 5. REF/Catalog No. H7493919312250, EMERGE MR, OUS 2.50mm X 12mm; 6. REF/Catalog No. H7493918912300, EMERGE MR US 3.00MM X 12MM; 7. REF/Catalog No. H7493919312300, EMERGE MR, OUS 3.00mm X 12mm; 8. REF/Catalog No. H7493918915150, EMERGE MR US 1.50MM X 15MM; 9. REF/Catalog No. H7493919315150, EMERGE MR, OUS 1.50mm X 15mm; 10. REF/Catalog No. H7493918915200, EMERGE MR US 2.00MM X 15MM, 11. REF/Catalog No. H7493919315200, EMERGE MR, OUS 2.00mm X 15mm; 12. REF/Catalog No. H7493919315250, EMERGE MR, OUS 2.50mm X 15mm; 13. REF/Catalog No. H7493918915300, EMERGE MR US 3.00MM X 15MM; 14. REF/Catalog No. H7493919315300, EMERGE MR, OUS 3.00mm X 15mm; 15. REF/Catalog No. H7493918915350, EMERGE MR US 3.50MM X 15MM; 16. REF/Catalog No. H7493919315350, EMERGE MR, OUS 3.50mm X 15mm; 17. REF/Catalog No. H7493919320200, EMERGE MR, OUS 2.00mm X 20mm; 18. REF/Catalog No. H7493919320250, EMERGE MR, OUS 2.50mm X 20mm; 19. REF/Catalog No. H7493919320300, EMERGE MR, OUS 3.00mm X 20mm; 20. REF/Catalog No. H7493919415120, EMERGE PUSH MR, OUS 1.20mm X 15mm.
Status
Terminated
Initiated
2019-10-24
Posted
not on file
Terminated
2023-03-20
Reason
Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.
Root cause
Process control
Action
The firm, Boston Scientific, sent an "Urgent Medical Device Removal" letter dated 10/28/2019 overnight mail delivery to its customers. The letter stated the issue, identified affected product and stated that product should not be used. The customers were instructed, if any product is identified to segregate immediately and return to Boston Scientific. The customers were also instructed to Immediately complete the RECALL REMOVAL REPLY VERIFICATION TRACKING FORM and Scan/e-mail to: BSCFieldActionCenter@bsci.com OR Fax to #: 1-866-213-1806. This form must be completed and returned in all cases even if you do not have any affected product. If you are a distributor, please note that the removal notification should be forwarded to your customers. If you have any questions, contact Boston Scientific Quality Systems at 763-494-1133 or email: BSCFieldActionCenter@bsci.com.
Quantity
14,145 devices total
Distribution
Nationwide including Washington D.C. and Puerto Rico. OUS to include Canada.
Lot, serial or model codes
Lot / GTIN Number / Use By: 1. 24112923 / 8714729799788 / 2022-03-12 2. Lot: 21362152 / 8714729806080 / 2020-07-06, 3. Lots: 21311605 / 8714729799894 / 2020-06-23, 23960557 / 8714729799894 / 2022-02-11, 23975925 / 8714729799894 / 2022-02-13, 24112926 / 8714729799894 / 2022-03-12.  4. Lots: 20524227 / 8714729806103 / 2019-12-31, 22402212 / 8714729806103 / 2021-03-15, 22909685 / 8714729806103 / 2020-12-06.  5. Lots: 23881230 / 8714729799917 / 2022-01-25, 23881291 / 8714729799917 / 2022-01-25, 24088571 / 8714729799917 / 2022-03-07, 24143068 / 8714729799917 / 2022-03-18.  6. Lots: 22374160 / 8714729806127 / 2021-03-09, 22532219 / 8714729806127 / 2021-04-16, 22760617 / 8714729806127 / 2021-02-05, 23021557 / 8714729806127 / 2021-01-11, 23181062 / 8714729806127 / 2021-09-10, 23578933 / 8714729806127 / 2021-04-16.  7. Lot: 22799195 / 8714729799931 / 2021-06-12.  8. Lots: 20361236 / 8714729806189 / 2019-10-31, 22619082 / 8714729806189 / 2021-05-03.  9. Lot: 24088574 / 8714729799993 / 2022-03-07.  10. Lots: 20985435 / 8714729806196 / 2020-04-04, 22416559 / 8714729806196 / 2021-03-19.  11. Lots: 21308852 / 8714729800002 / 2020-06-23, 21309387 / 8714729800002 / 2020-06-23, 21363021 / 8714729800002 / 2020-07-06, 23989173 / 8714729800002 / 2022-02-15.  12. Lots: 20909603 / 8714729800026 / 2020-03-15, 22734757 / 8714729800026 / 2021-05-29, 23047327 / 8714729800026 / 2021-08-03, 23946181 / 8714729800026 / 222-02-07, 24151134 / 8714729800026 / 2022-03-19.  13. Lot: 23924000 / 8714729806233 / 2022-02-04.  14. Lots: 23272475 /8714729800040 / 2021-09-27 , 23541254 / 8714729800040 / 2021-11-20.  15. Lot: 23901505 / 8714729806257 / 2022-01-30.  16. Lot: 24133473 / 8714729800064 / 2022-03-15.  17. Lots: 23272477 / 8714729800118 / 2021-09-21, 23324727 / 8714729800118 / 2021-10-08.  18. Lots: 20452512 / 8714729800132 / 2019-11-30, 20669615 / 8714729800132 / 2020-01-15, 23408005 / 8714729800132 / 2021-10-24.  19. Lots: 21500615 / 8714729800156 / 2020-08-08, 23518523 / 8714729800156 / 2021-11-15.  20. Lot: 23364688 / 8714729799702 / 2021-10-16.
510(k) numbers
["K113220","K121196","K130391","K163174"]
PMA numbers
not on file
Product code
LOX
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register