FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
- Recall number
- Z-0576-2015
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
- Status
- Terminated
- Initiated
- 2014-11-11
- Posted
- 2014-12-15
- Terminated
- 2015-02-24
- Reason
- Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
- Root cause
- Component design/selection
- Action
- Siemens sent a Safety Advisory notices letter dated November 11, 2014 to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions call 1-888-826-9702.
- Quantity
- 149
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- model numbers: 10592457, 10863171, 10863172, 10863173, 7502029 , 10496279, with multiple serial numbers: 1187, 3182, 3183, 3212, 3206, 3207, 1361, 10870, 12068, 10733, 3052, 2323, 1496, 3246, 2254, 1089, 1213, 14898, 2626, 1624, 2365, 3087, 13504, 11795, 13921, 2971, 12271, 13090, 9594, 2890, 1372, 12789, 3062, 1856, 2500, 13591, 14005, 3017, 3010, 1373, 9792, 13727, 1374, 14292, 2562, 3165, 2098, 2698, 3211, 2769, 1382, 1165, 2310, 13480, 2324, 2751, 2901, 2897, 2981, 3093, 3050, 3085, 14328, 100544, 100566, 100198, 100600, 100259, 100135, 100133, 100031, 100084, 100233, 200308, 200309, 200310, 200311, 200312, 200313, 200314, 200315, 200316, 200317, 200318, 200319, 200320, 200321, 200322, 200323, 200324, 100104, 100105, 100106, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100119, 100120, 100245, 100314, 200277, 100139, 100177, 100027, 100601, 100170, 100025, 100174, 100033, 100186, 100189, 100190, 100191, 100192, 100195, 100196, 100306, 100232, 100185, 100596, 100256, 100599, 100193, 100257, 100198, 100226, 100198, 100181, 100029, 100253, 100301, 100132, 100247, 100248, 100139, 100215, 100230, 100146, 100212, 100539.
- 510(k) numbers
- ["K082430","K093612","K132532"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28