Skip to the content

FDA Medical Device Recalls (openFDA)

Nova Biomedical Corporation: Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942

Recall number
Z-0576-06
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942
Status
Terminated
Initiated
2006-01-27
Posted
2006-02-28
Terminated
2006-10-19
Reason
Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs
Root cause
Other
Action
Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.
Quantity
498 units
Distribution
Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI
Lot, serial or model codes
All S/N''s
510(k) numbers
["K020297"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1219029
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register