FDA Medical Device Recalls (openFDA)

INO Therapeutics (dba Ikaria): INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0., delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and comprehensive alarm systems. The primary clinical settings are Neonatal Intensive Care Units (NICU) and to transport of neonates.

Recall number
Z-0575-2015
Recalling firm
INO Therapeutics (dba Ikaria)
Product
INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0., delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and comprehensive alarm systems. The primary clinical settings are Neonatal Intensive Care Units (NICU) and to transport of neonates.
Status
Terminated
Initiated
2014-10-17
Posted
2014-12-12
Terminated
2015-04-16
Reason
An issue has been identified in the INOmax DSIR system that could result in monitored Nitric Oxide (NO) concentration reporting lower than expected. This issue only pertains to those devices manufactured using a specific version of the Monitoring Circuit Board.
Root cause
Process change control
Action
Ikaria contacted consignees by phone beginning 10/17/2014. A follow-up letter was sent to customers beginning 10/22/2014. The letter described the issue, requested that affected devices not be used and to return to Ikaria. Customers with questions can contact Ikaria Customer Care at 1-877-566-9466.
Quantity
9
Distribution
US Distribution in states of: VA and PA only.
Lot, serial or model codes
Serial Numbers: DS20080152, DS20090925, DS20101076, DS20101499, DS20110886, DS20110965, DS20120099, DS20120150, DS20120169.
510(k) numbers
["K131686"]
PMA numbers
not on file
Product code
MRN
Firm FEI
3004531588
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register