Sterilmed, Inc.: Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Component: N/A
- Recall number
- Z-0573-2026
- Recalling firm
- Sterilmed, Inc.
- Product
- Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Component: N/A
- Status
- Open, Classified
- Initiated
- 2025-10-08
- Posted
- 2025-11-20
- Terminated
- not on file
- Reason
- Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
- Root cause
- Reprocessing Controls
- Action
- Initial phone call notification was performed 10/10/2025 to affected customers. Customers were instructed to immediately stop use of identified devices. Affected products should be segregated. Follow-up written notification was sent to customers on 10/17/25 which included instructions on how to return affected products.
- Quantity
- 6
- Distribution
- United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah
- Lot, serial or model codes
- Lot Code: GTIN/UDI: 10888551004725 LOT# SERIAL #: 2196112 999253 2200877 937157 2219499 1182771 2219926 1048031 2220881 1126584 2234882 2304200759
- 510(k) numbers
- ["K043453"]
- PMA numbers
- not on file
- Product code
- OWQ
- Firm FEI
- 3011276773
- Firm city
- Plymouth
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as