Skip to the content

FDA Medical Device Recalls (openFDA)

Clark Laboratories, Inc. (dba,Trinity Biotech USA): Captia Syphilis-G Test Kit Product Usage: This is an enzyme immunoassay for the qualitative detection of IgG antibodies to T. pallidum in serum specimens, to be used in conjunction with non-treponemal testing to provide serological evidence of infection with T. pallidum ( the agent of syphilis). It is also intended for testing of serum or plasma specimens to screen blood and/or plasma donors to exclude a history of syphilis.

Recall number
Z-0572-2014
Recalling firm
Clark Laboratories, Inc. (dba,Trinity Biotech USA)
Product
Captia Syphilis-G Test Kit Product Usage: This is an enzyme immunoassay for the qualitative detection of IgG antibodies to T. pallidum in serum specimens, to be used in conjunction with non-treponemal testing to provide serological evidence of infection with T. pallidum ( the agent of syphilis). It is also intended for testing of serum or plasma specimens to screen blood and/or plasma donors to exclude a history of syphilis.
Status
Terminated
Initiated
2013-08-27
Posted
2013-12-26
Terminated
2016-05-24
Reason
Trinity Biotech recalled Trinity Biotech Captia Syphilis IgG kits due to false negative results.
Root cause
Process design
Action
Trinity Biotech issued an Urgent Product Information Request letter to their customer on August 27, 2013. The letter identified the affected product, problem and actions to be taken. The request was to determine if their customer did or did not backfill microtiter-strips during the analytical process. If a customer backfilled, they were advised to return the completed request form. On September 11, 2013 Trinity Biotech issued a notice to their customers advising them to re-evaluate any negative patient results for those who perform backfilling. For questions contact Trinity Biotech Help Desk at 800-325-3424, option 2.
Quantity
38 kits
Distribution
USA Nationwide Distribution in the state of WA
Lot, serial or model codes
Product Code: 800-960, Lot number: 082, Model: 10 - plate kit
510(k) numbers
not on file
PMA numbers
not on file
Product code
LIP
Firm FEI
1318354
Firm city
Jamestown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register