Skip to the content

FDA Medical Device Recalls (openFDA)

Sterilmed Inc: Tourniquet Cuffs Model #'s: INSDT-7024-02 INSDT-7218-01 SMI553322 SMI7010-7002 SMI7010-7006 SMI7010-7007 SMI7010-7008 SMI7010-7010 SMI7010-7011 SMI7010-7012 SMI7010-7015 STR5921-018-136 STR5921-018-235 STR5921-024-135 STR5921-024-136 STR5921-024-146 STR5921-024-235 STR5921-030-135 STR5921-030-235 STR5921-034-135 STR5921-034-136 STR5921-034-235 STR5921-044-135 STR5921-044-136 STR5921-044-235 STR5921-212-135 ZIM60-7070-02 ZIM60-7070-03 ZIM60-7070-04 ZIM60-7070-05 ZIM60-7070-06 ZIM60-7070-07 ZIM60-7070-101 ZIM60-7070-102 ZIM60-7070-103 ZIM60-7070-104 ZIM60-7070-105 ZIM60-7070-106 ZIM60-7070-107 ZIM60-7075-006 ZIM60-7075-03 ZIM60-7075-04 ZIM60-7075-05 ZIM60-7075-06 ZIM60-7075-07 ZIM60-7075-101 ZIM60-7075-102 ZIM60-7075-103 ZIM60-7075-104 ZIM60-7075-106 ZIM60-7075-107 ZIM60-7080-102 and ZIM60-7080-103 A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff.

Recall number
Z-0572-2009
Recalling firm
Sterilmed Inc
Product
Tourniquet Cuffs Model #'s: INSDT-7024-02 INSDT-7218-01 SMI553322 SMI7010-7002 SMI7010-7006 SMI7010-7007 SMI7010-7008 SMI7010-7010 SMI7010-7011 SMI7010-7012 SMI7010-7015 STR5921-018-136 STR5921-018-235 STR5921-024-135 STR5921-024-136 STR5921-024-146 STR5921-024-235 STR5921-030-135 STR5921-030-235 STR5921-034-135 STR5921-034-136 STR5921-034-235 STR5921-044-135 STR5921-044-136 STR5921-044-235 STR5921-212-135 ZIM60-7070-02 ZIM60-7070-03 ZIM60-7070-04 ZIM60-7070-05 ZIM60-7070-06 ZIM60-7070-07 ZIM60-7070-101 ZIM60-7070-102 ZIM60-7070-103 ZIM60-7070-104 ZIM60-7070-105 ZIM60-7070-106 ZIM60-7070-107 ZIM60-7075-006 ZIM60-7075-03 ZIM60-7075-04 ZIM60-7075-05 ZIM60-7075-06 ZIM60-7075-07 ZIM60-7075-101 ZIM60-7075-102 ZIM60-7075-103 ZIM60-7075-104 ZIM60-7075-106 ZIM60-7075-107 ZIM60-7080-102 and ZIM60-7080-103 A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff.
Status
Terminated
Initiated
2008-08-08
Posted
2009-01-13
Terminated
2011-04-26
Reason
SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Root cause
Packaging process control
Action
Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
Quantity
1565
Distribution
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
Lot, serial or model codes
n/a
510(k) numbers
not on file
PMA numbers
not on file
Product code
KCY
Firm FEI
3001237663
Firm city
Osseo
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register