FDA Medical Device Recalls (openFDA)
Data Innovations, Inc.: Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
- Recall number
- Z-0571-2014
- Recalling firm
- Data Innovations, Inc.
- Product
- Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
- Status
- Terminated
- Initiated
- 2013-11-06
- Posted
- 2013-12-24
- Terminated
- 2014-12-05
- Reason
- When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
- Root cause
- Software design
- Action
- Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.
- Quantity
- 24 systems
- Distribution
- Nationwide Canada and Britain
- Lot, serial or model codes
- Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 3001451449
- Firm city
- South Burlington
- Firm state
- VT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28