Skip to the content

FDA Medical Device Recalls (openFDA)

Securitas Healthcare LLC: Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application

Recall number
Z-0570-2025
Recalling firm
Securitas Healthcare LLC
Product
Arial Server Software Versions 11.1.4 through 11.1.6, SKU 54305, in conjunction with the Arial Mobile Application or Arial Care Giver Application
Status
Open, Classified
Initiated
2024-11-01
Posted
2024-11-27
Terminated
not on file
Reason
Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
Root cause
Software Design Change
Action
An URGENT: MEDICAL DEVICE CORRECTION letter dated 10/31/24 was sent to customers. Necessary Actions to be Taken by Clients Instructions for determining which versions of the Arial Server Software you are using are attached hereto as Exhibit C. Clients using Arial Server Software versions 11.1.4.9, 11.1.5.5, or 11.1.6.2 should: 1. Not decline previously cleared calls through the Arial App; 2. Restart the Arial Server Software if an end user has declined a previously completed call using the Arial App. After restarting the Arial Server Software, calls from the affected call point(s) will begin to be received by the Arial Server Software and the Arial Server Software will notify end users as intended. Instructions for restarting the Arial Server Software are included on Exhibit B. 3. Contact Securitas Healthcare Technical Support at 1 (800) 380-8883 to upgrade the Arial Server Software to version 11.1.7 or later. For general questions concerning this communication please contact Securitas Healthcare Technical Support at 1 (800) 380-8883. We ask that you please complete and return a copy of the acknowledgement form attached hereto as Exhibit A.
Quantity
111 systems
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Switzerland.
Lot, serial or model codes
All Systems using Arial Server Software versions 11.1.4 through 11.1.6 in conjunction with the Arial Mobile Application or Arial Care Giver Application
510(k) numbers
not on file
PMA numbers
not on file
Product code
IQA
Firm FEI
1929691
Firm city
Lincoln
Firm state
NE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register