FDA Medical Device Recalls (openFDA)
Becton Dickinson & Co.: BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated Microbiology system is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria.
- Recall number
- Z-0570-2013
- Recalling firm
- Becton Dickinson & Co.
- Product
- BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated Microbiology system is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria.
- Status
- Terminated
- Initiated
- 2012-11-06
- Posted
- 2012-12-19
- Terminated
- 2013-01-09
- Reason
- An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, BD, sent an "URGENT PRODUCT RECALL" letter dated November 2012 via the telephone and fax to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and discard any remaining packages for replacement and to complete and return the Acknowledgement Receipt Form, whether or not you have any inventory remaining, via fax to: BD Regulatory Compliance at 410-316-4258. BD will issue replacements for the discarded materials. If further assistance is needed regarding replacements please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department at 1-800-638-8663.
- Quantity
- 12,250 tests
- Distribution
- Nationwide distribution: USA including states of: AR, CA, KY, OK, RI, WI, and WY.
- Lot, serial or model codes
- Lot number 2053373
- 510(k) numbers
- ["K070809"]
- PMA numbers
- not on file
- Product code
- LON
- Firm FEI
- 1119779
- Firm city
- Sparks
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28