FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF MLE 24-012. Product packaged in boxes of 6 pouches. Box of 6 has correct box label of ACMI USA Elite System and USA Series Roller Bar Electrode 24/26 Fr (8/8.667 mm), Model: REF RB
- Recall number
- Z-0568-2024
- Recalling firm
- Olympus Corporation of the Americas
- Product
- ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF MLE 24-012. Product packaged in boxes of 6 pouches. Box of 6 has correct box label of ACMI USA Elite System and USA Series Roller Bar Electrode 24/26 Fr (8/8.667 mm), Model: REF RB,
- Status
- Open, Classified
- Initiated
- 2023-10-02
- Posted
- 2023-12-15
- Terminated
- not on file
- Reason
- Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.
- Root cause
- Labeling Change Control
- Action
- An Urgent Medical Device Recall notice, issued via UPS Ground October 27, 2023, advises customers of the mislabeled product pouches. Customers are to examine and quarantine any affected inventory, contact Olympus customer service to arrange for return/replacement, and forward notice to any facility where product was further distributed. Acknowledge receipt of the recall notice through the Olympus recall portal at https://olympusamerica.com/recall (enter recall number "0432"). Customers requiring additional information may contact Cynthia.Ow@olympus.com.
- Quantity
- 162 pieces (27 boxes, 6 per box)
- Distribution
- US: AL, FL, GA, KS, LA, MD, MI, NC, NJ, NY, TX, WI & WV and Canada.
- Lot, serial or model codes
- UDI-DI: 00821925036413 (Incorrect Right Angle Cutting Loop pouch label) UDI-DI: 00821925036444 (Correct on Roller Bar Electrode box label) Lot: KR260855
- 510(k) numbers
- ["K890328"]
- PMA numbers
- not on file
- Product code
- FAS
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28