FDA Medical Device Recalls (openFDA)

Olympus Corporation of the Americas: ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF MLE 24-012. Product packaged in boxes of 6 pouches. Box of 6 has correct box label of ACMI USA Elite System and USA Series Roller Bar Electrode 24/26 Fr (8/8.667 mm), Model: REF RB

Recall number
Z-0568-2024
Recalling firm
Olympus Corporation of the Americas
Product
ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF MLE 24-012. Product packaged in boxes of 6 pouches. Box of 6 has correct box label of ACMI USA Elite System and USA Series Roller Bar Electrode 24/26 Fr (8/8.667 mm), Model: REF RB,
Status
Open, Classified
Initiated
2023-10-02
Posted
2023-12-15
Terminated
not on file
Reason
Incorrect pouch labels, the contents within the mislabeled pouch were Roller Bar electrodes.
Root cause
Labeling Change Control
Action
An Urgent Medical Device Recall notice, issued via UPS Ground October 27, 2023, advises customers of the mislabeled product pouches. Customers are to examine and quarantine any affected inventory, contact Olympus customer service to arrange for return/replacement, and forward notice to any facility where product was further distributed. Acknowledge receipt of the recall notice through the Olympus recall portal at https://olympusamerica.com/recall (enter recall number "0432"). Customers requiring additional information may contact Cynthia.Ow@olympus.com.
Quantity
162 pieces (27 boxes, 6 per box)
Distribution
US: AL, FL, GA, KS, LA, MD, MI, NC, NJ, NY, TX, WI & WV and Canada.
Lot, serial or model codes
UDI-DI: 00821925036413 (Incorrect Right Angle Cutting Loop pouch label) UDI-DI: 00821925036444 (Correct on Roller Bar Electrode box label) Lot: KR260855
510(k) numbers
["K890328"]
PMA numbers
not on file
Product code
FAS
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register