Skip to the content

FDA Medical Device Recalls (openFDA)

Galt Medical Corporation: Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-002-06 Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-006-16

Recall number
Z-0568-2021
Recalling firm
Galt Medical Corporation
Product
Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-002-06 Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-006-16
Status
Terminated
Initiated
2020-11-16
Posted
2020-12-14
Terminated
2022-11-07
Reason
Due to a manufacturing issue, the catheter tip geometry is incorrect (thicker wall) resulting in a more obtuse tip angle.
Root cause
Process control
Action
On 11/16/2020, the firm sent an "URGENT-Medical Recall Notice" to customers to inform them that certain Lots of catheters may have a tip geometry with a thicker wall, resulting in a more obtuse tip angle. This obtuse tip angle, if present in the catheter tip, can cause increased insertion resistance through the epidermis layer of the skin. The Recalling Firm is instructing customers to: Note- An updated Customer Notification Letter was sent on 11/20/2020 via email. 1. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users. 2. Complete the enclosed Field Correction Reply Forms and email or fax them to the attention of Corrections Coordinator at quality@galtmedical.com or 214-778-1433. The Recalling Firm has enclosed one form to account for inventory at the customer's facility and an additional form for capturing the same details from their end-users. The form lists the product number, lot number and quantity that the customer has received. The Recalling Firm is pointing out that It is important that even if a customer does not have any product remaining in their possession that they fill out the attached forms (Attachment A & B) noting zero quantity to be returned and email or fax the forms to the Recalling Firm. 3. Ship the returned product to GALT MEDICAL CORP. using Galts carrier account information listed on the forms. 4. Reference Return Authorization Number RMA# GALT_2004 on the outside of the shipping box and include a copy of the Field Correction Reply Forms with your shipment. 5. Once the completed Field Correction Reply Forms have been received and processed, the Recalling Firm will issue a credit to the customer for the product returned. Customer are asked to place a new PO notating Replacement Parts for new products. The cause of the problem has been identified and corrective actions have been implemented to assure this problem does not reoccur. If you h
Quantity
9,920 catheters (updated on 03/17/2021)
Distribution
U.S.: FL, GA, MA, NC, NJ, OH, PA, Puerto Rico and UT. O.U.S: Canada, Netherlands, Turkey, United Kingdom, and UAE
Lot, serial or model codes
Catalog Number: DRC-002-06  Lot Numbers/UDI Code: 20108509/(01)0084 1268 100398(17)240428(10)20108509 20139676/(01)0084 1268 100398(11)200520(17)240519(10)20139676 20148911/(01)0084 1268 100398(11)200609(17)240608(10)20148911 20160253/(01)0084 1268 100398(11)200707(17)240706(10)20160253 20189100/(01)0084 1268 100398(11)200821(17)240820(10)20189100  Updated Information provided 03/16/2021: For Catalog Number: DRC-002-06 additional Lot Numbers: -20080351 -20234573 -20297230 -21008452  Additional Products: Catalog Number: DRC-006-16 Lot Numbers: -20217235 -20217236 -20217237
510(k) numbers
not on file
PMA numbers
not on file
Product code
KGZ
Firm FEI
3000718472
Firm city
Garland
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register