FDA Medical Device Recalls (openFDA)
Medline Industries Inc: Medline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including a waste bag, for use by the Interventional Radiology Department; Packaged for Medline Industries, Inc.; Reorder #DYNJ35993D This custom pack includes components necessary for a specific procedure.
- Recall number
- Z-0568-2013
- Recalling firm
- Medline Industries Inc
- Product
- Medline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including a waste bag, for use by the Interventional Radiology Department; Packaged for Medline Industries, Inc.; Reorder #DYNJ35993D This custom pack includes components necessary for a specific procedure.
- Status
- Terminated
- Initiated
- 2012-11-08
- Posted
- 2012-12-19
- Terminated
- 2012-12-20
- Reason
- The non-sterile Waste Bag component in the custom pack was attached to the outside of the pack in its own packaging. The product was not labeled or identified as sterile, but was assumed to be sterile by the customer.
- Root cause
- Packaging
- Action
- The firm, Medline Industries, Inc., telephoned the consignee/customer on November 8, 2012. The phone call informed the customer of the product, problem and actions to be taken. The customer was instructed to inform the staff at the facility not to use the pack until it had been reworked to replace the non-sterile waste bag with the sterile equivalent. It was also noted that the product does not need to be returned as it is being replaced directly at the facility. If you have any question call 847-837-2759 or email lsimmons@medline.com.
- Quantity
- 105 packs
- Distribution
- Nationwide distribution: New York only.
- Lot, serial or model codes
- Reorder #DYNJ35993D, lot numbers 12HD2126, exp. 5/31/2015; 12IB5014, exp. 7/31/2015 and 12JB8311, exp. 5/31/2015
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 1417592
- Firm city
- Mundelein
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28