Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries Inc: Medline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including a waste bag, for use by the Interventional Radiology Department; Packaged for Medline Industries, Inc.; Reorder #DYNJ35993D This custom pack includes components necessary for a specific procedure.

Recall number
Z-0568-2013
Recalling firm
Medline Industries Inc
Product
Medline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including a waste bag, for use by the Interventional Radiology Department; Packaged for Medline Industries, Inc.; Reorder #DYNJ35993D This custom pack includes components necessary for a specific procedure.
Status
Terminated
Initiated
2012-11-08
Posted
2012-12-19
Terminated
2012-12-20
Reason
The non-sterile Waste Bag component in the custom pack was attached to the outside of the pack in its own packaging. The product was not labeled or identified as sterile, but was assumed to be sterile by the customer.
Root cause
Packaging
Action
The firm, Medline Industries, Inc., telephoned the consignee/customer on November 8, 2012. The phone call informed the customer of the product, problem and actions to be taken. The customer was instructed to inform the staff at the facility not to use the pack until it had been reworked to replace the non-sterile waste bag with the sterile equivalent. It was also noted that the product does not need to be returned as it is being replaced directly at the facility. If you have any question call 847-837-2759 or email lsimmons@medline.com.
Quantity
105 packs
Distribution
Nationwide distribution: New York only.
Lot, serial or model codes
Reorder #DYNJ35993D, lot numbers 12HD2126, exp. 5/31/2015; 12IB5014, exp. 7/31/2015 and 12JB8311, exp. 5/31/2015
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
1417592
Firm city
Mundelein
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register