FDA Medical Device Recalls (openFDA)

Pfm Medical Inc: Veta Peritoneal Catheter Kit, Curled, Ref No. M7225 Product Usage: Veta Peritoneal Dialysis Catheters are intended for acute and chronic access to the peritoneal cavity.

Recall number
Z-0567-2015
Recalling firm
Pfm Medical Inc
Product
Veta Peritoneal Catheter Kit, Curled, Ref No. M7225 Product Usage: Veta Peritoneal Dialysis Catheters are intended for acute and chronic access to the peritoneal cavity.
Status
Terminated
Initiated
2014-10-20
Posted
2014-12-12
Terminated
2015-04-23
Reason
PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.
Root cause
Process control
Action
PFm Medical sent an Urgent Device Recall Notification letter dated November 3, 2014 to all their customers who purchased the catheters and other medical devices. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the recall acknowledgement/reply form and fax it to (760) 758-1167 or email it to customerservice@pfmmedicalusa.com. Customers with any questions are instructed to contact Jessica Greene at (760) 758-8749.
Quantity
85 units
Distribution
Worldwide Distribution - US Nationwide in the states of MA, NC, UT, PA, TX, OH, LA and worldwide to: Germany and Malaysia.
Lot, serial or model codes
Lot No. 1212-005
510(k) numbers
not on file
PMA numbers
not on file
Product code
GCB
Firm FEI
3013666218
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register