FDA Medical Device Recalls (openFDA)
Pfm Medical Inc: ASEPT Peritoneal Drainage System, Ref No. P09080002 Product Usage: The pfm medical ASEPT Peritoneal Drainage System is a tunneled, indwelling catheter used to drain accumulated fluid from the abdomen.
- Recall number
- Z-0565-2015
- Recalling firm
- Pfm Medical Inc
- Product
- ASEPT Peritoneal Drainage System, Ref No. P09080002 Product Usage: The pfm medical ASEPT Peritoneal Drainage System is a tunneled, indwelling catheter used to drain accumulated fluid from the abdomen.
- Status
- Terminated
- Initiated
- 2014-10-20
- Posted
- 2014-12-12
- Terminated
- 2015-04-23
- Reason
- PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.
- Root cause
- Process control
- Action
- PFm Medical sent an Urgent Device Recall Notification letter dated November 3, 2014 to all their customers who purchased the catheters and other medical devices. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the recall acknowledgement/reply form and fax it to (760) 758-1167 or email it to customerservice@pfmmedicalusa.com. Customers with any questions are instructed to contact Jessica Greene at (760) 758-8749.
- Quantity
- 83 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of MA, NC, UT, PA, TX, OH, LA and worldwide to: Germany and Malaysia.
- Lot, serial or model codes
- Lot No. 1211-002, 1210-016
- 510(k) numbers
- ["K083873"]
- PMA numbers
- not on file
- Product code
- LJS
- Firm FEI
- 3013666218
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28