Biomet 3i, LLC: Radiographic Transparency for Certain and External Connection Tapered Implants. Images that assist with measurements for dental implants.

Recall number
Z-0564-2016
Recalling firm
Biomet 3i, LLC
Product
Radiographic Transparency for Certain and External Connection Tapered Implants. Images that assist with measurements for dental implants.
Status
Terminated
Initiated
2015-11-16
Posted
2015-12-31
Terminated
2016-12-14
Reason
Images contained within radiographic transparency ART 1036 are undersized.
Root cause
Software design (manufacturing process)
Action
BIOMET3i sent an" Urgent Medical Device Recall letter " dated November, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Responsibilities: 1. Please review this notice and check your inventory for the affected item by comparing any on-hand ART 1036 Transparencies to a standard 8.5" x 11" sheet of paper. The non-conforming ART 1036 is approximately 7.8" x 10.2". 2. Immediately quarantine and remove all affected items from service. 3. Complete the attached Business Reply Form and either: a. fax it to 561-514-6316 or b. email it to postmarket@biomet.com or c. return it along with the affected product using the included shipping label 4. Return affected product to BIOMET 3i. For assistance or other questions that you may have relative to this notice, please contact BIOMET 3i at 1- 800-342-5454 or 1-561-776-6700. Team members are available to assist you 8:00am to 6:00pm (Eastern), Monday through Friday.
Quantity
686
Distribution
Worldwide Distribution - US Distribution including Puerto Rico., and to the states of : CA, CO, FL, IL, MA, MI, MO, NY, OH, SC, TX, VA, WI, WV and Hawaii., and to the countries of : Australia, Germany, France, Greece, Northern Mariana Islands, Taiwan and Uruguay.
Lot, serial or model codes
Model Number: ART 1036, Rev. C 03/13; Purchase Order # 184427.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDP
Firm FEI
1038806
Firm city
Palm Beach Gardens
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register