Baxter Healthcare Corp.: HomeChoice Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R

Recall number
Z-0564-03
Recalling firm
Baxter Healthcare Corp.
Product
HomeChoice Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R
Status
Terminated
Initiated
2002-12-23
Posted
2003-02-21
Terminated
2007-12-11
Reason
Potential for low ultrafiltrate volumes when using the Low Fill Mode during Continuous Cycling Peritoneal Dialysis (CCPD) Therapy
Root cause
Other
Action
Safety Alert letters dated 12/23/02 were mailed to the medical directos of facilities identified as having patients who use the low fill volume cassette with the HomeChoice machines. The letters informed them of the potential low ultrafiltrate volumes and requested that they review the patient list provided with the letter and verify that all of their patients who use the device in the low fill mode are included.
Quantity
not on file
Distribution
Nationwide, Guatemala, Chile, Brazil, Turkey, Singapore, Taiwan, Canada, Belgium, Czech Republic, Israel, Italy, Jordan, Lebanon, Malta, Spain, Sweden, Turkey and the United Kingdom.
Lot, serial or model codes
All HomeChoice Systems with software versions 8.51 and 8.52.
510(k) numbers
["K012988"]
PMA numbers
not on file
Product code
FKX
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register