Skip to the content

FDA Medical Device Recalls (openFDA)

Spacelabs Healthcare, Ltd.: Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.

Recall number
Z-0563-2021
Recalling firm
Spacelabs Healthcare, Ltd.
Product
Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is easy to install and to use.
Status
Terminated
Initiated
2020-11-06
Posted
not on file
Terminated
2024-10-18
Reason
During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.
Root cause
Software design
Action
On 11/06/20, Urgent Medical Device Correction notices were mailed to customers. Customers were asked to do the following: Any comments associated with ABP, Resting, or Rhythm tests that existed in your system prior to the date of your upgrade should be reviewed for accuracy before being used to inform any clinical decision. Weigh the benefits versus the risks when deciding whether to use the comments associated with ABP, Resting, or Rhythm tests which existed in your system prior to the date of your upgrade of Sentinel. We recommend that you brief your staff regarding the issue listed above. Complete and return the customer reply form or use the following link to acknowledge receipt of this letter: https://www.spacelabshealthcare.com/support/recall-reply-form/ For additional information or technical assistance, please contact: Global Technical Support at, +44 (0) 1992 507740, or gtsdc@spacelabs.com
Quantity
600
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countries of AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BELGIUM, BOLIVIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, Republic of Macedonia, SAUDI ARABIA, Serbia, SINGAPORE, Slovakia, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.
Lot, serial or model codes
Systems that have been upgraded from Version 9 or below to Version 10 or above.
510(k) numbers
["K152881"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3002807992
Firm city
Hertford
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register