Trumpf Medical Systems, Inc.: Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5

Recall number
Z-0563-2016
Recalling firm
Trumpf Medical Systems, Inc.
Product
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
Status
Completed
Initiated
2015-11-02
Posted
not on file
Terminated
not on file
Reason
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Root cause
Other
Action
TRUMPF sent an " Urgent Field Safety Information" letter dated December 8, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken by the customer: 1. Perform an inspection of the lighting systems (see Appendix 1 for instructions). If any of the symptoms listed below are observed, immediately remove the light from operation and contact your Trumpf Medical Service Representative or designee. Signs that indicate a possible fall of the spring arm: The movement of the light and spring arm is difficult for the user, The lights have stopped being operational, and/or A gap is noticed between the spring arm and the central axis (see the attached Appendix 1 for an example). 2. Follow the User Manual: No unauthorized servicing of the spring arm is permitted. Snap rings are one use only and replacements must be purchased from your authorized Trumpf Medical partner. If you have any questions concerning this request or the procedure as outlined, please contact TRUMPF Medical Systems, Inc. at 888-474-9359
Quantity
20,713
Distribution
USA (nationwide) Distribution to the states of : AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KY, ME, MA, MN, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WY., and Internationally to Canada.
Lot, serial or model codes
4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo; 4038310 - TruLight trio/quad; 1565068 - iLED single (TUSM); 1565160 - iLED duo (TUSM); 1565161 - iLED trio/quad (TUSM); 1574759 - TruLight single (TUSM); 1574850 - TruLight duo (TUSM); 1574851 - TruLight trio/quad (TUSM)
510(k) numbers
["K061317","K102758"]
PMA numbers
not on file
Product code
FSY
Firm FEI
3003184737
Firm city
Charleston
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register