Skip to the content

FDA Medical Device Recalls (openFDA)

Cygnus LLC: Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

Recall number
Z-0563-06
Recalling firm
Cygnus LLC
Product
Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.
Status
Terminated
Initiated
2005-02-04
Posted
2006-02-28
Terminated
2007-02-01
Reason
The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
Root cause
Other
Action
A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
Quantity
20 units
Distribution
All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741
Lot, serial or model codes
Serial Numbers:  For MTF: -- 034135 through 034144; For Cascade -- 034119 through 034128.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQC
Firm FEI
1000121662
Firm city
Paterson
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register