Skip to the content

FDA Medical Device Recalls (openFDA)

Ziehm Imaging Inc: Ziehm Vista/Vista Plus C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic Imaging

Recall number
Z-0562-2014
Recalling firm
Ziehm Imaging Inc
Product
Ziehm Vista/Vista Plus C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic Imaging
Status
Terminated
Initiated
2013-11-21
Posted
2013-12-24
Terminated
2014-06-23
Reason
Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
Root cause
Device Design
Action
Ziehm Imaging sent an "URGENT RECALL NOTICE" letter dated November 22, 2013, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customers. Customers were advised and instructed to: Your Model and Serial number can be found on the product label attached to the side of the mobile C-arm Stand. Our records indicate that your facility has one or more of the products listed above that are being used in your facility. Recall of the product was initiated because according to our manufacturing documentation, the referenced device installed at your facility was not equipped with an emergency stopping device for vertical travel of the c-arm imaging assembly. To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device identification and label. The issue: According to our manufacturing documentation, the referenced device installed at your facility was not equipped with an emergency stopping device. To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device. The manufacturer Ziehm Imaging GmbH advised customers they may continue to operate the affected device with caution until the correction has been carried out on-site. During this time Ziehm Imaging GmbH advises the operator to be aware of the following safety measures which may resolve the possibility of potentially hazardous situations by: " Operating the power-off switch located either on the C-arm Stand top panel or on the monitor cart " Removing the power cable from the wall outlet or, " Activate the push-buttons for reverse vertical movement and keeping the pushbutton pressed until patient /operator is no longer in the risk zone or the device stops Please alert all staff members working with the device to these potential hazardous situations, and familiarize them with all possibilities to resolve the hazardous situation
Quantity
83 devices
Distribution
USA (Nationwide) Distribution including the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MT, MA, MD, MI, MN, MN, MO,MS, NC ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Hawaii.
Lot, serial or model codes
5652 5704 5758 5772 5792 5840 5653 5714 5759 5773 5796 5844 5654 5715 5760 5774 5798 5846 5655 5716 5761 5775 5805 5847 5656 5717 5762 5776 5809 5854 5695 5718 5763 5777 5810 5855 5696 5720 5764 5779 5814 5860 5697 5721 5765 5780 5816 5861 5698 5722 5766 5781 5823 5862 5699 5723 5767 5782 5828 5863 5700 5749 5768 5783 5830 5864 5701 5750 5769 5784 5833 5865 5702 5751 5770 5785 5838 5866 5703 5755 5771 5787 5839
510(k) numbers
["K851031"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2027299
Firm city
Orlando
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register