Terumo Cardiovascular Systems Corporation: Sarns 8000 Roller pump for Sarns System 8000 Heart Lung Machine console, 115V, Catalog No 16402, Terumo Cardiovascular System Corp Ann Arbor, MI. The system 8000 is indicated for use in extracorporeal circulation of blood for arterial perfusion, reginal perfusion and cardiopulmonary bypass procedures.

Recall number
Z-0562-2011
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns 8000 Roller pump for Sarns System 8000 Heart Lung Machine console, 115V, Catalog No 16402, Terumo Cardiovascular System Corp Ann Arbor, MI. The system 8000 is indicated for use in extracorporeal circulation of blood for arterial perfusion, reginal perfusion and cardiopulmonary bypass procedures.
Status
Terminated
Initiated
2010-11-02
Posted
2010-12-08
Terminated
2011-10-19
Reason
The system 8000 roller pumps may fail resulting in a persistent pump stop.
Root cause
Other
Action
The firm sent "URGENT MEDICAL DEVICE CORRECTION" dated 11/02/2010 to customers. The letter describes the product, problem and corrective actions taken and/or needed to be taken by the customers. The letter states the system responds to a failed inductor by displaying an overspeed message and stopping the roller pump without an audible alert. The pump cannot be restarted. The customers/users are instructed to stop the pump, replace the pump with a backup or if necessary use the handcrank. The letter states that Terumo will replace the roller pump at the next preventative maintenance visit - or sooner- if the customer has a service contract. If the customer does not have a service contract, Terumo will contact the customer within 2 weeks to set up an appointment for the service. Terumo CVS expects all affected inductors to be replaced to be replaced within three months. Call Terumo CVS Customer Service upon completion of the case at 800-521-2818. The customer was also instructed to complete and return the enclosed CUSTOMER RESPONSE FORM Please note that we are not recommending that you stop using the roller pumps for your Sarns system 8000. For any questions, call toll free 800-262-3304 or phone 734-663-4145.
Quantity
95 roller pumps.
Distribution
Worldwide distribution: USA including states of CA, FL, NC and NY and countries of United Arab Emirates, China, Ecuador, and Panama.
Lot, serial or model codes
lots 19945-19954 and 19985-20001.
510(k) numbers
["K915183"]
PMA numbers
not on file
Product code
JOR
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register