Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: Philips HeartStart Vehicle Wall Mount, Model M5528A with Philips HeartStart DC Power Module, Model M5529A. The HeartStart MRx portable external defibrillator has optional accessories which include a wall mount and a DC power module. The wall mount is intended for storage and transport of MRx devices in land based transport vehicles. The power module is an additional optional accessory intended to be connected to the wall mount as a means for charging MRx devices (batteries) when stored in the wall mount. The DC power Module is also intended for use with an extension power cable, which is exclusively intended to be used to connect the DC module directly to the MRx. Philips private labels and distributes these devices which are designed and manufactured by Laerdal Medical.

Recall number
Z-0562-2010
Recalling firm
Philips Healthcare Inc.
Product
Philips HeartStart Vehicle Wall Mount, Model M5528A with Philips HeartStart DC Power Module, Model M5529A. The HeartStart MRx portable external defibrillator has optional accessories which include a wall mount and a DC power module. The wall mount is intended for storage and transport of MRx devices in land based transport vehicles. The power module is an additional optional accessory intended to be connected to the wall mount as a means for charging MRx devices (batteries) when stored in the wall mount. The DC power Module is also intended for use with an extension power cable, which is exclusively intended to be used to connect the DC module directly to the MRx. Philips private labels and distributes these devices which are designed and manufactured by Laerdal Medical.
Status
Terminated
Initiated
2009-11-12
Posted
2009-12-23
Terminated
2012-01-25
Reason
HeartStart Vehicle Wall Mount (Model M5528A) with DC Power Module (Model M5529A)--Incorrect wiring during installation by user may result in a smoke or fire hazard.
Root cause
Device Design
Action
Philips Healthcare, Inc. notified customers by an "URGENT - Voluntary Medical Device Correction" letter, plus revised Instructions For Use (IFU) for the Vehicle Wall Mount accessory and the DC Power Module accessory of the HeartStart MRx which includes a revised description of correct installation and verification of performance. Customers were asked to follow the Actions in the customer/user section of the Urgent - Voluntary Medical Device Correction Notification. In this section, customers were given the instructions to correct their devices. Distribution of the Urgent - Voluntary Medical Device Correction Notification along with the IFU will be managed by Philips representatives in each affected geography. Direct questions or concerns about the recall to your local Philips representative.
Quantity
3,265 units
Distribution
Class II Recall -- Worldwide Distribution -- United States, Canada, Australia, Brazil, China, France, Germany, Hong Kong, Iceland, India, Indonesia, Italy, Japan, Jordan, Kenya, Libyan Arab Jamahiriya, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Romania, Saudi Arabia, Spain, Sri Lanka, Switzerland, Trinidad, Tobago, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Units distributed by Philips Healthcare Inc. and shipped between September 1, 2004 and October 1, 2009.
510(k) numbers
["K051134"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register