Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Emergency Response Systems, Inc.: LIFEPAK 500 AED.

Recall number
Z-0562-05
Recalling firm
Medtronic Emergency Response Systems, Inc.
Product
LIFEPAK 500 AED.
Status
Terminated
Initiated
2005-02-03
Posted
2005-03-10
Terminated
2006-08-21
Reason
Potential to not detect patients at the low end of the patient impedance range.
Root cause
Device Design
Action
On 2/3/05 the firm issued a 'Medical Device Correction' letter dated January 2005 to their customers. The letter advised customers of the issue and stated a representative from the firm would visit each facility to update affected unit. On 2/25/05 a nationwide press release was issued. On 3/1/05 the firm began telephoning each customer, this was followed by another letter titled MEDICAL DEVICE RECALL. The letter advises of the issue and recommends defibrillators remain in service when no alternative is available until the unit can be corrected. Each unit will be replaced or corrected. On 4/22/05 the firm issued another letter, via certified mail return receipt, expanding their recall. Letter was issued to all customers affected by the expansion of the recall. On 4/22/05 the firm issued a press release advising of the expansion of the recall. The firm's web site also provides a listing of affected units.
Quantity
2,349
Distribution
Devices were distributed to 750 domestic consignees dn 16 international consignees.
Lot, serial or model codes
Serial numbers - various, noncontiguous Part numbers - 3005400-XXX, D3005400-XXX, U3005400-XXX
510(k) numbers
not on file
PMA numbers
not on file
Product code
MKJ
Firm FEI
3015876
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register