Skip to the content

FDA Medical Device Recalls (openFDA)

Helena Laboratories, Inc.: Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) FOR EXPORT ONLY

Recall number
Z-0561-2019
Recalling firm
Helena Laboratories, Inc.
Product
Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) FOR EXPORT ONLY
Status
Terminated
Initiated
2017-02-24
Posted
not on file
Terminated
2023-10-20
Reason
Due to some cards in this lot reported to have experienced variations in results, (verified by wave form), resulting in some cards with shorter or longer times than expected.
Root cause
Device Design
Action
On 02/24/2017, the firm sent an email to affected users, with a Withdrawal Notification. The firm informed the customers/distributors that Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected. The firm instructed the customers/distributors to disposes of affected lot, as a preventive quality measure. On 04/12/2017, the firm mailed formal letters to all customers. PUBLIC CONTACT: "Name: Dr. Jessica Jones Hanka "Title: Technical Services Manager "Address: 1530 Lindbergh Drive, Beaumont, TX 77707 "Telephone #: (409) 842-3714 x1177 "Email: jhanka@helena.com
Quantity
27 kits
Distribution
US: Products under control of Manufacturer in TX OUS: Four International Distributors - Australia, Netherlands, Italy, and England
Lot, serial or model codes
Lot Codes: 1-16-5701 Model/Catalog Number: 5701
510(k) numbers
not on file
PMA numbers
not on file
Product code
JPB
Firm FEI
1618982
Firm city
Beaumont
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register