Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien LLC: Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2

Recall number
Z-0561-2018
Recalling firm
Covidien LLC
Product
Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2
Status
Open, Classified
Initiated
2017-08-07
Posted
not on file
Terminated
not on file
Reason
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Root cause
Device Design
Action
All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
Quantity
not on file
Distribution
Nationwide including PR, Canada, China
Lot, serial or model codes
UDI 10884521706606, Lot codes: S6MG007PX,S6MG010PX,S6MG016PX,S7AG002PX,S7BG005X,S7BG009X,S7DG004X,S7FG006X
510(k) numbers
["K133821"]
PMA numbers
not on file
Product code
NEY
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register