Inpeco S.A.: Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. Hardware versions: FLX-247-00 Firmware versions: COBASa_2.2.0, COBASa_2.3.0

Recall number
Z-0560-2024
Recalling firm
Inpeco S.A.
Product
Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. Hardware versions: FLX-247-00 Firmware versions: COBASa_2.2.0, COBASa_2.3.0
Status
Open, Classified
Initiated
2023-11-03
Posted
2023-12-15
Terminated
not on file
Reason
Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
Root cause
Software Design Change
Action
Inpeco issued Urgent Medical Device Correction Letter (Recall-identifier: 3010825766 11/06/23 002 C) via email on 11/3/23. Letter states reason for recall, health risk and action to take: Please apply the following instructions to detect the sample tubes potentially impacted by the issue (Step 1) and to manage them according to the specific occurred scenario (Step 2). If Step 1 leads to determine that the issue has not occurred, Step 2 is not to be performed. Note: The test orders status displayed on the IUI is not overwritten neither changed in any way due to the issue: it would be possible to monitor the turn-around time of sample tubes processed by the CO8 IM in order to identify any overdue sample tube with tests still to be executed. Contact your local technical support provider if you need assistance in implementing these actions. Your service provider will contact you to schedule the firmware upgrade. Until the service visit please maintain awareness on this notice and apply the actions recommended above. Transmission of this Recall Letter: Please transfer this notice to whom it might concern. Please complete and return the Customer Letter Receipt Confirmation and Implementation Check form attached to this letter within 30 days directly to the email address specified in the email communication. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: regulatory.affairs@inpeco.com Phone: (+41) 91 9118 224
Quantity
1 unit US: 14 units OUS
Distribution
Worldwide - US Nationwide distribution in the state of MA and the countries of AUSTRIA, BELGIUM, DENMARK, ITALY, NORWAY, SPAIN.
Lot, serial or model codes
FlexLab UDI-DI:07640172340004 Hardware versions: FLX-247-00 Firmware versions: COBASa_2.2.0, COBASa_2.3.0 Serial Number US: FLX.0208 Serial Numbers OUS: Serial Number FLX.0116 FLX.0165 FLX.0178 FLX.067 FLX.0127 FLX.0128 FLX.0150 FLX.0154 FLX.012 FLX.0121 FLX.0132 FLX.031 FLX.048 FLX.049
510(k) numbers
["K121012"]
PMA numbers
not on file
Product code
CEM
Firm FEI
3010825766
Firm city
Lugano
Firm state
not on file
Firm country
Switzerland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register