Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion 5 M20

Recall number
Z-0559-2024
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion 5 M20
Status
Open, Classified
Initiated
2023-11-03
Posted
2023-12-15
Terminated
not on file
Reason
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Root cause
Device Design
Action
Philips issued Urgent Medical Device Correction (UMDC) letter US customers in the US on 11/3/23 and a Field Safety Notice (FSN) to customers outside the US. Letter states reason for recall, health risk and action to take: As a remedy, Philips is replacing the existing covers of all the affected Philips Allura and Philips Azurion with the monoplane fixed ceiling mounted systems with a redesigned L-arm rotation cover (bolted instead of spring clips) to make sure it is resistant to external collisions. Philips will contact all affected customers to arrange for a Field Service Engineer visit to replace the Larm rotation cover free of charge (reference FCO72200510 for the Philips Allura systems including the Philips Poly G H5000, FCO72200511 for the Philips Azurion systems). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
3821 units
Distribution
Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Lot, serial or model codes
Mdoel/Product Name/UDI-DI: 722078 Azurion 7 M12 (01)00884838085251(21); 722079 Azurion 7 M20 (01)00884838085268(21); 722223 Azurion 7 M12 (01)00884838099241(21); 722224 Azurion 7 M20 (01)00884838099258(21); 722227 Azurion 5 M12 (01)00884838099227(21); 722228 Azurion 5 M20 (01)00884838099234(21).
510(k) numbers
["K200917"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register