PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)

Recall number
Z-0558-2024
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)
Status
Open, Classified
Initiated
2023-11-03
Posted
2023-12-15
Terminated
not on file
Reason
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Root cause
Device Design
Action
Philips issued Urgent Medical Device Correction (UMDC) letter US customers in the US on 11/3/23 and a Field Safety Notice (FSN) to customers outside the US. Letter states reason for recall, health risk and action to take: As a remedy, Philips is replacing the existing covers of all the affected Philips Allura and Philips Azurion with the monoplane fixed ceiling mounted systems with a redesigned L-arm rotation cover (bolted instead of spring clips) to make sure it is resistant to external collisions. Philips will contact all affected customers to arrange for a Field Service Engineer visit to replace the Larm rotation cover free of charge (reference FCO72200510 for the Philips Allura systems including the Philips Poly G H5000, FCO72200511 for the Philips Azurion systems). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
9991 units
Distribution
Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Lot, serial or model codes
Model/Name/UDI-DI: 722001 Allura Xper FD10C; 722003 Allura Xper FD10; 722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table; 722016 Integris H5000 C / Allura 9C ; 722018 Integris Allura 9; - 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane).
510(k) numbers
["K161563","K162859","K200917"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register