Siemens Medical Solutions USA, Inc: Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Recall number
Z-0558-2019
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Status
Terminated
Initiated
2018-10-26
Posted
not on file
Terminated
2020-05-12
Reason
This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.
Root cause
Process control
Action
On October 26, 2018, the firm, Siemens Healthineers, distributed "URGENT: MEDICAL DEVICE CORRECTION" letters to affected customers. The letters informed customers of the product, problem and actions to be taken. The customers were instructed to do the following: promptly notify and instruct accordingly all the staff at your organization who need to be aware of this notice and Immediately examine your inventory and quarantine product subject to recall (if applicable). In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall.If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner and inform us of the identity of the device's new owner where possible. The letter also stated that Siemens will replace the coolant pump on affected lots of cooling units to correct the issue. Siemens' service organization will contact customers to arrange a date to perform the corrective action with Update Instruction AX013/18/S. To request an earlier appointment, feel free to contact the service organization at 1-800-888-7436. If you have any questions, contact Regulatory Technical Specialist at 610-219-4834 or email to: anastasia.sokolova@siemens-healthineers.com.
Quantity
64 (16 in US)
Distribution
US Distribution to states of: AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, and UT.
Lot, serial or model codes
Serial nos: 910045-09 911354-02 910045-19 908883-23 910045-15 910045-14 910045-25 908883-04 911354-25 908883-18 908883-11 910045-21 911354-26 908883-06 908883-02 911354-27
510(k) numbers
["K073290"]
PMA numbers
not on file
Product code
IZI
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register