Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer/CAS Power Cord, Sesamoid Plasty, NA, Zimmer/CAS, Montreal (Quebec), Canada; REF 20-8000-070-12. The device is the power cord component for the Sesamoid Plasty CAS workstation which connects the workstation to the power mains.

Recall number
Z-0558-2010
Recalling firm
Zimmer Inc.
Product
Zimmer/CAS Power Cord, Sesamoid Plasty, NA, Zimmer/CAS, Montreal (Quebec), Canada; REF 20-8000-070-12. The device is the power cord component for the Sesamoid Plasty CAS workstation which connects the workstation to the power mains.
Status
Terminated
Initiated
2009-11-02
Posted
2009-12-22
Terminated
2010-10-15
Reason
The power cord female receptacle is not recessed sufficiently and may present a shock hazard.
Root cause
Device Design
Action
Zimmer sales staff were notified by letter dated 11/2/09 and instructed to locate the units and to upgrade the cords and to notify consignees by copy of a letter addressed to risk managers and dated 11/2/09.
Quantity
71
Distribution
Nationwide.
Lot, serial or model codes
Workstation serial numbers SP014, SP015, SP019 through SP022, SP025 through SP033, SP035, SP038, SP039, SP043, SP044, SP047, SP049, SP052, SP057, SP058, SP060, SP061, SP062, SP064, SP065, SP069, SP072, SP074 through SP083, SP089, SP090, SP091, SP093, SP094, SP098, SP099, SP121, SP124, SP125, SP127, SP128, SP131 through SP135, SP151 through SP154, SP157, SP175, SP176, SP183, SP184, SP185, SP190 and SP198.
510(k) numbers
["K060336","K071714","K071929"]
PMA numbers
not on file
Product code
HAW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register