Skip to the content

FDA Medical Device Recalls (openFDA)

Conmed Corporation: FRAZIER Atraumatic Soft Tip Surgical Suction Instrument with Control Vent and Obturator, 10 French, Product Code: 0031100, 20/C. STERILE, EO, Rx ONLY. MANUFACTURED FOR CONMED CORPORATION, 310 Broad Street, Utica, NY 13501

Recall number
Z-0558-2008
Recalling firm
Conmed Corporation
Product
FRAZIER Atraumatic Soft Tip Surgical Suction Instrument with Control Vent and Obturator, 10 French, Product Code: 0031100, 20/C. STERILE, EO, Rx ONLY. MANUFACTURED FOR CONMED CORPORATION, 310 Broad Street, Utica, NY 13501
Status
Terminated
Initiated
2007-09-12
Posted
2008-02-02
Terminated
2012-08-30
Reason
Sterility (package integrity) compromised -- Frazier and Poole suction instruments may have an inadequate seal. The product pouch seal made during assembly of this product, may be incomplete or not present.
Root cause
Packaging process control
Action
URGENT-Device Recall Letters (dated 9/10/07) and FIELD CORRECTIVE ACTION CERTIFICATION forms were sent via Certified Mail to affected customers the week of 9/09/07 (mailed on 9/12/07). The recall letters informed the consignees of the recall due to potentially inadequate seals of the instruments, and instructed them to review their inventory of the instruments checking them for inadequate seals. The letters also instructed the consignees to complete the form and fax it back by 9/21/07. For those consignees that have any of the recalled instruments or may have questions, please contact ConMed Quality Assurance at 1-800-552-0138 ext. 5225 between the hours of 7 am - 6 pm (MST). International customers were notified on 9/18/07. A second recall letter (dated 10/09/07) was sent to non-respondents of the initial mailing. ConMed plans to send a third recall letter to the non-responders during the week of 12/10/07 which addresses sub-recall notifications to the distributors' customers.
Quantity
19,334 instruments
Distribution
Nationwide and Worldwide Distribution
Lot, serial or model codes
Product Code: 0031100; All lot codes manufactured between 10/04/06 and 7/31/07 (lots 061004 through 070731). Lot codes on boxes and packaging contain a 6 digit lot code: For example, lot 061004: the first 2 digits represent the year manufactured (06 represents 2006), next 2 digits represent the month (10 represents October), the next 2 digits represent the day of the month (04).
510(k) numbers
not on file
PMA numbers
not on file
Product code
GCJ
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.