FDA Medical Device Recalls (openFDA)
Covidien LLC: Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm (reinforced), Material CA15L2
- Recall number
- Z-0557-2018
- Recalling firm
- Covidien LLC
- Product
- Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm (reinforced), Material CA15L2
- Status
- Open, Classified
- Initiated
- 2017-08-07
- Posted
- not on file
- Terminated
- not on file
- Reason
- The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
- Root cause
- Device Design
- Action
- All consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
- Quantity
- not on file
- Distribution
- Nationwide including PR, Canada, China
- Lot, serial or model codes
- UDI 10884521706583, Lot codes: S6MG005PX,S6MG008PX,S6MG012PX,S6MG019PX,S7AG001PX,S7AG006X,S7AG010X,S7AG013X,S7BG001X,S7BG003X,S7BG007X,S7BG013X,S7CG002X,S7CG005X,S7CG007X,S7CG013X,S7EG009X,S7FG003X,S7FG004X,S7FG007X
- 510(k) numbers
- ["K133821"]
- PMA numbers
- not on file
- Product code
- NEY
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28