Skip to the content

FDA Medical Device Recalls (openFDA)

Shimadzu Medical Systems: Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.

Recall number
Z-0554-2021
Recalling firm
Shimadzu Medical Systems
Product
Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.
Status
Open, Classified
Initiated
2020-11-18
Posted
not on file
Terminated
not on file
Reason
The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.
Root cause
Software design
Action
On 11/18/2020 Shimadzu sent an "Urgent Medical Device Recall Notice" to all affected consignees. In addition to informing consignees about the recall the firm provided the following information. 1. Shimadzu has released a mandatory Service Bulletin on applicable products for the update of the positioner software. The update of the software and the testing of the positioner movement should be performed by your Authorized Shimadzu Service provider. 2. Your authorized Distributor has acknowledged receipt the Service Bulletin (XR-202026) and the recall Customer Letter. They have been instructed to provide you a copy of this letter so that you may understand the scope of the recall. Necessary software for this recall and instruction to complete have been sent to your Distributer and they should be scheduling time to perform the corrective actions. 3. Once the corrective work has been completed your Distributor will report to Shimadzu that the work has been completed so that we may track and inform the appropriate Regulatory Agencies of its completion. Actions to be taken by customer: A. Please check the list of systems affected by this recall. If your site is on the applicable list, please pay attention to the following points until the corrective work has been completed. . (1) This event does not occur in any position other than the Left or Right diagonal position, so please avoid using this product in diagonal positions. (2) When using the product in the Left or Right diagonal position, please use the direct memory function and encourage C-arm operation without using the lever. (3) If the C-arm starts to operate unintentionally, press the red "Stop Switch" button immediately. B. If you feel there is a problem with the c-arm movement, please discontinue use of the system and contact your local service provider. C. Please complete the Medical device Recall Return Response form and submit to drodriguez@shimadzu-usa.com D. Shimadzu
Quantity
7 units
Distribution
US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.
Lot, serial or model codes
Serial Numbers: 40AD58C55001, 40AD58C4B001, 40AD58C24001, 40AD58522001, 40AD58C1C001, 101X968101, 091X955901
510(k) numbers
["K062360"]
PMA numbers
not on file
Product code
IZI
Firm FEI
1000125362
Firm city
Torrance
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register