Skip to the content

FDA Medical Device Recalls (openFDA)

The Binding Site Group, Ltd.: Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of rheumatoid factor in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings

Recall number
Z-0553-2018
Recalling firm
The Binding Site Group, Ltd.
Product
Rheumatoid Factor (RF) Kit for use on SPAPLUS, Product #LK151.S.A Intended for the quantitative in-vitro measurement of rheumatoid factor in serum using the Binding Site SPAPLUS analyser. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis. This test should be used in conjunction with other laboratory and clinical findings
Status
Terminated
Initiated
2018-01-04
Posted
not on file
Terminated
2020-06-09
Reason
The low and high controls for the product are returning results above the assigned values.
Root cause
Under Investigation by firm
Action
Binding Site sent an Urgent Field Safety Notice dated December 22, 2017. Advise on action to be taken by the User: " The new QC values detailed below should be used for the remainder of the shelf life of this Rheumatoid Factor Kit for use on SPAPLUS. "Input the revised QC values into your SPAPLUS instrument "Sign and return the TSWS18 E-Back Form accompanying this notification within 5 business days "Refer to Section 9 of the Product Insert SIN282(.A) if QCs are out of range following reassignment. For further questions, please call (858) 453-9177.
Quantity
49 kits to the U.S. office (32 kits further distributed)
Distribution
US Distribution was made to CA. There was no foreign/military/government distribution.
Lot, serial or model codes
Lot #416595, Exp. 31/03/2018; UDI #05051700018319
510(k) numbers
["K160070"]
PMA numbers
not on file
Product code
DHR
Firm FEI
3002808340
Firm city
Birmingham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register