Skip to the content

FDA Medical Device Recalls (openFDA)

Moving Solutions, Inc.: Faaborg patient lift/hoist, battery operated lifts. The Solution/Nordic series lifts have electric V-shape base widening, with a weight capacity of 210 lbs. to 365 lbs. depending on the base model. Please note: the Faaborg Person Lift arm assembly can be interchanged on the various lift series bases. The arm assembly is labeled ''Faaborgliften Made in Denmark'', while the base plate reads in part ''Faaborg Rehab Technic Lift Model: VL-***** Serial No: **** Capacity: *** KG. Manufactured: **-**** 5600 Faaborg Phone+4570205553 Fax+4570205554'' A sticker is also present on the base reading ''Moving Solutions 800-228-7980 Fax 630-963-9588''. The Users Guide refers to the lift as a ''Faaborghoist''.

Recall number
Z-0553-04
Recalling firm
Moving Solutions, Inc.
Product
Faaborg patient lift/hoist, battery operated lifts. The Solution/Nordic series lifts have electric V-shape base widening, with a weight capacity of 210 lbs. to 365 lbs. depending on the base model. Please note: the Faaborg Person Lift arm assembly can be interchanged on the various lift series bases. The arm assembly is labeled ''Faaborgliften Made in Denmark'', while the base plate reads in part ''Faaborg Rehab Technic Lift Model: VL-***** Serial No: **** Capacity: *** KG. Manufactured: **-**** 5600 Faaborg Phone+4570205553 Fax+4570205554'' A sticker is also present on the base reading ''Moving Solutions 800-228-7980 Fax 630-963-9588''. The Users Guide refers to the lift as a ''Faaborghoist''.
Status
Terminated
Initiated
2004-01-21
Posted
2004-07-20
Terminated
2007-10-22
Reason
Excessive wear of the hanger bar bolt, that connects the lift arm to the sling spreader bar, may cause the bolt to snap allowing the patient to fall.
Root cause
Other
Action
Letters were sent to all Faaborg lift customers on 11/29/01 and 01/24/04 advising the users that all bolts in connection with the hanger bar must be inspected, greased and tightened, and all defective or worn down bolts and washers must be changed. The 1/24/04 letters included an upgraded maintenance schedule and an Important Service/Safety Reminder, as well as a supply of hanger bar bolt washers. Any questions were directed to Moving Solutions' Service Department at 800-228-7980. Follow-up letters were sent to the accounts on 3/11/04, referencing the 1/21/04 letter and instructing the accounts not to use the lifts until the bolts have been inspected and retrofitted with the washers. Copies of the revised maintenance schedule and operating instructions were included.
Quantity
204 lifts
Distribution
Nationwide to nursing homes, residential schools and home patients, and internationally to Canada.
Lot, serial or model codes
all Solution/Nordic series models, all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
1000305914
Firm city
Downers Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register