FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness of breath and hyperinflation Model Number: SVS-V9-00
- Recall number
- Z-0551-2022
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Spiration Valve System (SVS) Valve in Cartridge -A one-way endobronchial valves indicated for adult patients with shortness of breath and hyperinflation Model Number: SVS-V9-00
- Status
- Open, Classified
- Initiated
- 2021-12-23
- Posted
- not on file
- Terminated
- not on file
- Reason
- Mislabeled Patient Charge Label indicating model number SVS-V7-00 (7mm) instead of SVS-V9-00 (9mm)
- Root cause
- Labeling Change Control
- Action
- Gyrus ACMI Inc. (Olympus) issued Urgent Medical Device Correction Letter dated 12/23/21. Letter states reason for recall, health risk and action to take: Inspect your inventory for the referenced device and identify any devices with the model number and lot number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Please contact Customer Service 1-888-524-7266, option 1, to obtain a Return Material Authorization. Olympus will issue a credit or replacement to your facility for your affected product. 2. Review your facility s use of the patient charge labels. If your facility has used these mislabeled patient charge labels in the patient s chart, or scanned into electronic medical records and/or patient implant cards please request replacement labels by indicating the quantity in the Recall Portal (see Step 3). Olympus will ship replacement labels for you to use in correcting your patient records. In the case of labels sent with patients at discharge for their personal records, include an explanation of the situation as described above. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number "0403 and provide your contact information as indicated in the portal. As you are most likely aware and as provided in the labeling for the product, for any subsequent procedures with that patient, the SVS endobronchial valve size selection is based on patient anatomy using a sizing kit (SVS-VSK-06) and the fit is verified by a trained clinician after placement according to the product s instruction manual. As a result, the mislabeled patient charge label was determined to not adversely affect medical treatment. However, Olympus recognizes that there may be confusion about what product size was previously used and is therefore sending this notice to you. Olympus requests you to report any complaints, including product l
- Quantity
- 35 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, MA, NJ, PA, SC, TX, WI.
- Lot, serial or model codes
- Lot Number: WS147887-01 UDI: 896506002309 Expiration Date: 4/30/2024
- 510(k) numbers
- not on file
- PMA numbers
- ["P180007"]
- Product code
- NJK
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28