Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm Diameter, 70 cm length, Item Number 47-2490-098-01 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.

Recall number
Z-0551-2018
Recalling firm
Zimmer Biomet, Inc.
Product
Zimmer Natural Nail System Tear Drop Guide Wire 2.4 mm Diameter, 70 cm length, Item Number 47-2490-098-01 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.
Status
Terminated
Initiated
2017-06-05
Posted
not on file
Terminated
2019-04-17
Reason
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
Root cause
Package design/selection
Action
On June 5, 2017, Zimmer Biomet distributed Urgent Medical Device Recall notices to their customers via FedEx and email. **Hospital Risk Managers are advised to:** 1. Review this notification and ensure affected team members are aware of the contents. 2. Complete the Certification of Acknowledgement portion of Attachment 1 a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 3. Assist your Zimmer Biomet sales representative quarantine all affected product. **Distributors are advised to:** 1. Review this notification and ensure affected personnel are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 4. Note that any hospitals that received direct shipments of this product from Zimmer Biomet will be sent a copy of the Risk Manager directly. It is important that you review the list of hospitals included with the email notification sent to your facility to identify additional accounts Zimmer Biomet has not notified. Using the Additional Accounts Form provided with the email notice sent to your facility, return contact information for any additional hospitals that may have received or used the affected product within the past 12 months. If there are no additional users to notify, please confirm this by annotating and returning the form provided indicating that no additional users have been identified. 5. Return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form to Zimmer Biomet within 2 weeks. a. For each return, send a copy of Attachment 1 to corporatequality.postmarket@zimmerbiomet.com. b. Include a hardcopy of Attachment 1 with your shipment for i
Quantity
1,631 units
Distribution
Nationwide within U.S.A and Puerto Rico; International: Australia, Brazil, Canada, China, Germany, India, Japan, Netherlands & Singapore
Lot, serial or model codes
All Lot Numbers Expiry Date Before March 2022
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register