Integra LifeSciences Corp. d.b.a. Integra Pain Management: Integra Cranial Access Kit, Model Number HITHSP10, Rx Only. Integra Cranial Access Kits (CRAK) allows for burr hole access to the subarachnoid space or the lateral ventricles of the brain. The kit is intended to be used with an external drainage and monitoring system in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of Cerebrospinal Fluid (CSF), and to monitor ICP. These kits contain all components and accessories required for burr hole access into the cranium.

Recall number
Z-0551-2014
Recalling firm
Integra LifeSciences Corp. d.b.a. Integra Pain Management
Product
Integra Cranial Access Kit, Model Number HITHSP10, Rx Only. Integra Cranial Access Kits (CRAK) allows for burr hole access to the subarachnoid space or the lateral ventricles of the brain. The kit is intended to be used with an external drainage and monitoring system in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of Cerebrospinal Fluid (CSF), and to monitor ICP. These kits contain all components and accessories required for burr hole access into the cranium.
Status
Terminated
Initiated
2013-11-26
Posted
2013-12-20
Terminated
2014-04-09
Reason
Integra LifeSciences is voluntarily recalling any unexpired lots of HITHSP10 Cranial Access Kits due to identified labeling discrepancies. kit contains DuraPrep and Monocryl sutures whose labeling does not align with the intended use and labeling of the Cranial Access Kit.
Root cause
Labeling mix-ups
Action
Integra sent an Urgent Voluntary Medical Device Recall letter dated November 26, 2013, to all affected consignees. The letter identified the product, the problem, the action to be taken by the consignee and included a Recall Acknowledgement and Return form. Consignees were instructed to review their inventory and determine if they have any affected product, stop using the affected product immediately and remove from service. Consignees were instructed to complete the attached form even if they do not have the affected product in their inventory and return by email or fax as indicated on the form. When the form is received Customer Service will send an RMA number and directions to return the product. Customers with questions were instructed to contact Customer Service at 1-855-532-1723 or their Integra sales representative at 909-201-4099. For questions regarding this recall call 609-936-6822.
Quantity
731
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot No. W1302120 Expires 2/1/2014  Lot No. W1305038 Expires 8/1/2015  Lot No. W1306137 Expires 9/1/2014
510(k) numbers
["K042359"]
PMA numbers
not on file
Product code
JXG
Firm FEI
1000138491
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register