FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter Amia Automated Peritoneal Dialysis (APD) Cycler Set, REF 5C5479

Recall number
Z-0550-2024
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Amia Automated Peritoneal Dialysis (APD) Cycler Set, REF 5C5479
Status
Open, Classified
Initiated
2023-11-15
Posted
2023-12-14
Terminated
not on file
Reason
There is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.
Root cause
Process control
Action
Baxter issued an Urgent Medical Device Recall notice to its consignees on 11/15/2023 by US Mail. The notice explained the issue, potential risk, and requested the following: Isolate and cease all use of the affected units pending return.
Quantity
339,936 devices
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
UDI/DI 0085412153186, Lot Numbers: H23F23028, H23G11054, H23F24018, H23G12029, H23F24034, H23G13100, H23F24067, H23G14074, H23F25049, H23G15048, H23F25056, H23G16038, H23F26047, H23G17051, H23F26054, H23G18067, H23F27037, H23G19115, H23F27045, H23G20055, H23F27102, H23G22069, H23F28035, H23G23067, H23F28043, H23G24107, H23F29066, H23G25047, H23F30064, H23G25096, H23G01030, H23G26094, H23G02046, H23G27100, H23G05015, H23G28082, H23G06047, H23G30054, H23G07052, H23G31110, H23G08050, H23H04032, H23G09058, H23H05039, H23G10072
510(k) numbers
["K151525"]
PMA numbers
not on file
Product code
FKX
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register